Published ahead of print on July 27, 2006, doi:10.1164/rccm.200603-344OC
© 2006 American Thoracic Society doi: 10.1164/rccm.200603-344OC
Electromagnetic Navigation Diagnostic BronchoscopyA Prospective StudyDepartment of Pulmonary, Allergy and Critical Care Medicine, and Department of Radiology, The Cleveland Clinic Foundation, Cleveland, Ohio; and Ankara University School of Medicine, Department of Chest Diseases, Ankara, Turkey Correspondence and requests for reprints should be addressed to Thomas R. Gildea, M.D., Associate Staff, Department of Pulmonary, Allergy and Critical Care Medicine, The Cleveland Clinic Foundation, Desk A-90, 9500 Euclid Avenue, Cleveland, OH 44195. E-mail: gildeat{at}ccf.org
Rationale: Electromagnetic navigation bronchoscopy using superDimension/Bronchus System is a novel method to increase diagnostic yield of peripheral and mediastinal lung lesions. Objectives: A prospective, open label, single-center, pilot study was conducted to determine the ability of electromagnetic navigation bronchoscopy to sample peripheral lung lesions and mediastinal lymph nodes with standard bronchoscopic instruments and demonstrate safety. Methods: Electromagnetic navigation bronchoscopy was performed using the superDimension/Bronchus system consisting of electromagnetic board, position sensor encapsulated in the tip of a steerable probe, extended working channel, and real-time reconstruction of previously acquired multiplanar computed tomography images. The final distance of the steerable probe to lesion, expected error based on the actual and virtual markers, and procedure yield was gathered. Measurements: 60 subjects were enrolled between December 2004 and September 2005. Mean navigation times were 7 ± 6 min and 2 ± 2 min for peripheral lesions and lymph nodes, respectively. The steerable probe tip was navigated to the target lung area in all cases. The mean peripheral lesions and lymph nodes size was 22.8 ± 12.6 mm and 28.1 ± 12.8 mm. Yield was determined by results obtained during the bronchoscopy per patient. Results: The yield/procedure was 74% and 100% for peripheral lesions and lymph nodes, respectively. A diagnosis was obtained in 80.3% of bronchoscopic procedures. A definitive diagnosis of lung malignancy was made in 74.4% of subjects. Pneumothorax occurred in two subjects. Conclusion: Electromagnetic navigation bronchoscopy is a safe method for sampling peripheral and mediastinal lesions with high diagnostic yield independent of lesion size and location.
Key Words: diagnostic bronchoscopy electromagnetic navigation mediastinal lymph node solitary pulmonary nodule transbronchial biopsy Diagnostic yield of flexible bronchoscopy is limited by its inability to guide biopsy instruments directly to the lesion. The diagnostic success rate is dependent on the size and location of the lesion (17). The diagnostic yield of flexible bronchoscopy is expected to be between 20 and 84%. For lesions less than 2 cm in diameter, the diagnostic yield of flexible bronchoscopy is 14% for peripheral lesions in the outer third of the chest and as high as 31% if in the proximal two-thirds (8). The diagnostic yield of flexible bronchoscopy for mediastinal lymph nodes using transbronchial needle aspiration (TBNA) is reported to be between 15 and 83% (9). The diagnostic yield of TBNA in staging of bronchogenic carcinoma is reported to be between 50 and 60% (10). Many adjuvant technologies have been proposed to guide the tissue sampling, such as endobronchial ultrasound (EBUS) and computed tomography fluoroscopy (1113). The critical coupling of real-time guidance and the ability to precisely direct a biopsy instrument is critical to procedural success. The introduction of electromagnetic navigation bronchoscopy (ENB) with a steerable instrument has been previously described for peripheral lesions (1417). The superDimension/bronchus system (superDimension Ltd, Hertzliya, Israel) is an image-guided localization system, which is designed to guide bronchoscopic tools to predetermined points within the bronchial tree. The device uses three separate technologies that are combined to enable navigation of dedicated tools within the lung in real time. The first component is the planning software, which converts digital imaging and communications in medicine standards (DICOM) images from a computed tomography (CT) scan into multiplanar images with three-dimensional reconstruction and virtual bronchoscopy of the airways. The second component is a steerable probe that contains a position sensor attached to an eight-way steerable instrument that has the ability to navigate turns in the endobronchial tree. The third component is an electromagnetic (EM) board, which is a field generator connected to a computer containing the planning data. The exact position of the steerable probe when placed within the EM field is depicted on the system monitor. Thus, ENB has the ability to guide bronchoscopy instruments to reach lung targets for transbronchial biopsy (TBBX), brushing, or TBNA procedures (1417). A prospective open-label, single-center study was performed to determine the ability of ENB to reach peripheral lung lesions and mediastinal lymph nodes. Some of the results of this study have been previously reported in the form of an abstract at the International Association for the Study of Lung Cancer Society (IASLC) meeting in Barcelona in July 2005 (18).
Sixty subjects were enrolled between December 2004 and September 2005. Inclusion and exclusion criteria are detailed in Table 1. All subjects were candidates for nonemergency bronchoscopy of a suspected lesion, and/or enlarged mediastinal lymph nodes. Patients were enrolled if they were referred to our center for presumed difficult bronchoscopy, prior nondiagnostic bronchoscopy or with lesions traditionally not reachable by routine bronchoscopy. All subjects provided informed consent approved by the Institutional Review Board and were followed up for at least 6 wk after the procedures. In cases with regression or persistence of the lesion, the follow-up time extended until the current submission of data (average time 10.5 mo).
Endpoints The primary endpoint of the study was to determine the feasibility of ENB in reaching lung targets as demonstrated by location of steerable probe tip displayed on SDBS screen. The secondary endpoints were to determine the ability of ENB to assist in obtaining tissue with TBBX, brush, and/or TBNA. This was determined as follows. If bronchoscopy with ENB including TBBX/brush yielded a positive diagnosis or if TBNA resulted in a positive diagnosis or benign lymphoid tissue; it was considered of diagnostic value (ENB success). If bronchoscopy with ENB including TBBX/brush/TBNA resulted in a plausible negative diagnosis and it is further supported by additional procedures (CT-fine needle aspiration, surgery, clinical follow-up with imaging, etc.); it was also considered of diagnostic value (ENB success). If TBBX/brush/TBNA with ENB resulted was nondiagnostic, and additional procedures leading to positive diagnosis, the case was considered of nondiagnostic value (ENB failure). The safety of ENB during the bronchoscopic procedures was accessed. Yield is reported for both bronchoscopic procedures with ENB and also by lesion. All procedures were done without ROSE (rapid on-site cytopathologic evaluation), as it is not routinely available in our institution.
Procedure
Registration, Navigation, and Biopsy Registration is the process by which the computer links the five to six virtual fiducial markers to the actual position in the patient (Figure 2). Upon registration completion, the average fiducial target registration error (AFTRE) score was given in millimeters (mm). The AFTRE is the radius of expected difference of the location of the tip of the steerable probe in the actual patient compared with where it is expected to be in the virtual patient. After registration, navigation is performed with simultaneous advancement of the steerable probe toward the target and directing steerable probe to the lesion (Figure 3). The closest distance between the steerable probe tip and the lesion center was recorded.
When navigation was completed, the steerable probe is removed, leaving the extendable working channel through which brushings and TBBXs or TBNA were performed. Biopsies were performed using a C-arm fluoroscopy unit. All instruments were visualized under fluoroscopy only after navigation was completed to confirm proper function and position of the bronchoscopic tools relative to the lesion and the pleura. Brush biopsies involved two to three passes, and four pieces of tissue were obtained by TBBX. TBNA usually was done with 2 to 4 passes of a combination of 19-G and 22-G needles depending on physician choice (9, 1922).
Statistical Analysis
Among 60 cases, the ENB procedure was completed on 58 subjects (23 women, 35 men; ages 4386 yr; mean 67.91 ± 9.3) In two subjects, it could not be completed due to a mechanical failure and lack of cooperation each in one instance. From a total of 58 procedures, 36 subjects had only peripheral lesions sampled, 9 subjects had only lymph nodes sampled, and 13 subjects had both peripheral lesions and lymph nodes sampled. In addition, there were subjects with more than one peripheral lesions and/or lymph nodes for which ENB was used during bronchoscopy. Total numbers of peripheral lesions and lymph nodes sampled were 56 and 31, respectively, with mean sizes of 22.8 ± 12.6 mm and 28.1 ± 12.8 mm, respectively. Two patients had ENB procedures that were nondiagnostic but did not complete follow up (one subject died before any additional procedures could be performed, another was lost to follow up). Therefore, a total of 56 cases, including 54 peripheral lesions and 31 lymph nodes, were included into final analysis.
Ability to Reach Target Areas/Registration Accuracy/Navigation Accuracy
Procedure Times
Diagnostic Yield Overall, 80.3% (45 out of 56) bronchoscopy with ENB procedures resulted in obtaining diagnostic tissue. Seventy-four percent (40 out of 54) of procedures performed for peripheral lesions and 100% (31 out of 31) of lymph nodes were successfully sampled by ENB yielding a definitive diagnosis or benign lymphoid tissue (Table 3). Of the 54 lesions, 31 (57%) were less than 2 cm in the greatest diameter. Diagnostic yield was not significantly different by lesion sizes (Table 4). Diagnostic yield was not significantly affected by AFTRE (p = 0.1701). In correlation analysis, no significant relationship between diagnosis and size or location of peripheral lesions or lymph nodes was found (p > 0.05).
Using ENB, malignancy was diagnosed in 32 of the 43 subjects (74.4%). The most common diagnosis was nonsmall cell lung cancer including adenocarcinoma (n = 12), squamous cell carcinoma (n = 7), nonsmall cell lung cancer (unspecified) (n = 4), and large cell carcinoma (n = 3), accounting for 60.5% of the diagnosed cancers. Four patients (9%) had small cell lung cancer, and two patients (5%) had other malignancies, such as hepatocellular carcinoma and metastatic colon cancer diagnosed by ENB. When ENB was nondiagnostic, eleven subjects went on to have other diagnostic techniques. Thoracotomy (n = 6), CT-fine needle aspiration (n = 3), mediastinoscopy (n = 1), and positron emission tomography (n = 1) were used to confirm malignant disease (Table 5).
The diagnosis of specific benign diseases and other benign conditions were made in eight and five cases, respectively, accounting for all benign cases studied by ENB. For five cases in which ENB was nondiagnostic, a benign diagnosis was presumed based on repeated CT-scan imaging showing either resolution of the lesion or continued lesion size stability over an average of 10.5 mo of follow-up (at the time of this submission). This was considered a clinically plausible endpoint when either regression or no enlargement of the lesion was found during follow-up imaging (Table 5).
Safety
The salient study findings are that ENB performed under conscious sedation, via the transnasal approach, using standard bronchoscopy biopsy instruments is useful for safely obtaining diagnostic tissue from mediastinal lymph nodes and peripheral lesions independent of size or location. The excellent yield of ENB is related to the implementation of the following three pieces of integrated technology: (1) multiplanar CT planning, (2) electromagnetic navigation, and (3) the steerable probe. The sole use of multiplanar CT planning has been shown to increase the yield of diagnostic bronchoscopy (23). The addition of virtual bronchoscopy to the planning of multiplanar CT allows for appropriate identification of the airway, and may be different than what is anticipated from axial CT review. Scrolling through CT imaging planes allows identification of the "bronchus sign," which may guide the bronchoscopist to select the most appropriate segment that may not have been readily evident on a standard CT scan. The navigation component of ENB uses CT data to create virtual lungs, and tracks the movement of the probe as it moves in the actual patient. The computer processor speed acquisition of these images is instantaneous. The electromagnetic field does not expose the patient to additional radiation that would otherwise be needed for real-time CT acquisition or that would be needed for CT-guided percutaneous biopsies (1517). The steerable probe and extendable working channel are central to the utility of the technology. The ability of turning this probe beyond the bronchoscope may be very helpful, even independent of navigation (16, 17). Biopsy of peripheral lesions with standard flexible bronchoscopy is a blind procedure, and generally beyond the reach of the bronchoscope and thus the directional capability (6, 7, 24, 25). Not only does the steerable probe via the extendable working channel direct biopsy instruments toward the target, but combined with the navigation computer it gives an estimate of distance and relative location from the tip to the center of the lesion. The use of ENB requires training and development of a learning curve. In the planning phase, the user must learn to interact with a software package that incorporates multiplanar CT imaging and navigate in a three-dimensional rendering of the airways. If smaller lesions have to be approached, planning may start from the lesion going toward the closest reference point in a retrograde fashion if a bronchus can be found (2628). In the registration and navigation phase, the user must learn to account for respiratory movements, work a foot pedal, and correlate actual anatomy with virtual anatomy to mark registration points (Figure 2). During navigation, after "wedging" the bronchoscope, the user must learn how to advance and manipulate the steerable probe while simultaneously watching each of the moving multiplanar CT images and the tip view panel (Figures 1 and 3). In ENB, the course of the sensor probe is followed on computer screen and if it deviates from the target, it must be retracted and advanced in a different direction. To avoid repeated insertion into the wrong bronchus, an electronic track is left behind on the screen after retraction. When sampling tissue, the user should be comfortable using all of the routinely available biopsy techniques. Each user may have a variable number of cases until they are comfortable with each task. Fluoroscopy was not used during navigation, as it would interfere with the electromagnetic field. Fluoroscopy is comparatively unreliable in indicating the position of the sensor probe (29). Fluoroscopy was used when biopsies were taken to confirm utility of the instruments to assure that the instruments did not dislodge the EWC. This finding was noted with TBBX forceps and TBNA, but less with the cytology brush (Figure 4). Since fluoroscopy was used in all cases we cannot asses the incremental benefit of its use.
Success was evaluated in several different aspects. Guidance of the steerable probe to within 1 cm of the target center was achieved in all cases but not confirmed by other imaging techniques. Fluoroscopic location was not adequate to identify all lesions and not systematically collected to analyze. Success as defined by the ability get a definitive diagnosis may be limited as much by the instruments as by the ability to reach the lesion. One limitation of this study is that the number of biopsy attempts and techniques are variable. The use of the peripheral needle was limited by its ability to make the turn through the EWC to navigate to upper lobe lesions. The ability to localize, but not obtain diagnostic tissue has also been seen with peripheral EBUS (28, 30). Better flexible instruments may need to be developed to further increase yield. The design of this pilot study did make direct comparisons of ENB to other techniques, yet historical comparison to other diagnostic bronchoscopy studies is possible. A recent study of flexible bronchoscopy with adjunct PET scanning used to diagnose lesions less than 3 cm in size showed that standard bronchoscopy with the use of uniplanar fluoroscopy yielded a positive diagnosis in 53% of cases (31). Herth and colleagues used peripheral EBUS for lesions less than 3 cm in size. Peripheral EBUS was able to localize 89% of lesions, but sampling only resulted in 70% with a diagnosis (30). A similar study comparing peripheral EBUS to routine flexible bronchoscopy showed that EBUS was a superior to bronchoscopy in that yield was approximately 71% for lesions of less than 2 cm in size compared with 23.3% by routine bronchoscopy alone. EBUS with a guided sheath showed yield between 66 and 77% (32, 33), and the addition of virtual bronchoscopy to EBUS has also been evaluated. EBUS visualization was accomplished in 80%, but a diagnosis was only made in 63.3%, and only 44% if the lesion was less than 2 cm in size (28). EBUS used for mediastinal adenopathy has also been well studied. Yasufuku and colleagues report in 96% adequate specimen sampling with the use of the EBUS puncture bronchoscope (34). A randomized trial comparing standard TBNA to catheter based EBUS-TBNA showed that in areas other than the subcarinal lymph node EBUS was superior (58% versus 84% p < 0.0001) in obtaining adequate tissue (35). Each of these studies used a different technology. The integrated ultrasound puncture scope uses real-time localization, but can only be used for mediastinal disease. The catheter-based EBUS does not provide a real-time localization, as it must be removed from the working channel for the needle biopsy to be used and does not have a steering mechanism to make turns through the peripheral airways. The variation in the practice of diagnostic bronchoscopy internationally makes it very difficult to draw comparisons about different technologies. Differences in patient selection, lesion sizes, use of rigid bronchoscopy, conscious sedation versus deep or general anesthesia, biopsy techniques, the number of biopsies obtained, the use of ROSE, and operator skill major may dramatically change utility of any particular technology in a given setting. ENB appears to be superior to standard bronchoscopy and equivalent if not superior to other advanced techniques by historical comparison. While PET scanning is relatively sensitive and specific; it does not provide a tissue diagnosis. Although not systematically collected, patients in this trial had biopsy-proven cancers in the absence of a PET scan compatible with malignancy. PET scanning itself cannot rule out cancer and gives no definitive guidance for therapy in the absence of a tissue diagnosis. The ACCP guidelines for the staging of nonsmall cell cancer suggest PET scanning be used, but still requires additional tissue sampling before considering surgical resection (36). CT-guided percutaneous biopsy has reported yield of 64 to 97% (37). In a recent publication of 846 CT guided percutaneous procedures, the diagnostic accuracy was 94%, but the average lesion size in the study was 3.02 cm and the complication rate was 30% (27% pneumothorax and 3% hemoptysis) (38). The lesion size was not associated with risk of pneumothorax, but others have documented an 11-fold higher risk of pneumothorax with lesions less than 2 cm in size compared with lesions of more than 4 cm. If the lesion was more than 2 cm from the pleura, the risk of pneumothorax increased by fourfold (39). In the current study, several patients were deemed to be unsuitable for CT-guided biopsy due to risk of pneumothorax or bleeding or because the lesion were deemed inaccessible (e.g., lesions behind the sternum, near the upper mediastinum, or in the aorto-pulmonary window). ROSE was not used in this study as it is not routinely available in our center. While it has proven to be of high utility and cost effective (40), others have suggested that it would be unnecessary with more advanced biopsy techniques (35). The utility of ENB with and without ROSE has not yet been studied. There were five patients in whom a definitive diagnosis was not obtained. These patients had low clinical probability of malignancy by established criteria (4143), and their follow-up showed no difference in the size of the lesion as of the submission of this paper (follow-up period of 10.5 mo in average). These cases were therefore defined as diagnostic and negative as long as the pathology showed some abnormality but nondiagnostic and not "normal lung tissue." The short 10.5-mo follow-up period is not adequate for a definitive benign diagnosis, but since the subjects were not sent for additional tissue sampling, this was considered a reasonable endpoint in this study. In two cases, sarcoidosis was the final diagnosis. In these cases the mediastinal lymph nodes were successfully sampled but there were no granulomas. The BAL in each patient did reveal a high level of lymphocytosis with high CD4/CD8 ratio. The treating physicans in these cases used the bronchoscopic data to begin therapy and required no additional diagnostic tissue. These cases were recorded as an ENB success (as benign lymphoid tissue was obtained in both), and also as being diagnostic. No adverse events were considered to be directly related to the EMB itself. The two cases of pneumothorax were treated with a small-bore chest tube and recovered without incident. These were most likely related to the TBBX and are similar to the reported rate of pneumothorax during routine bronchoscopy of 4% (44, 45). There were no serious adverse events related to ENB. To our best knowledge, this is the first study in a large scale of patients with peripheral lesions and/or mediastinal lymph node enlargement to assess the utility of ENB. It is a safe procedure that, with some training, is associated with high diagnostic successhigher then those reported for routine diagnostic bronchoscopy and perhaps similar or superior to some other advanced techniques such as EBUS independent of size and location. Prospective comparison with other complementary methods such as EBUS may be studied in the future. We have discovered limitations of the currently available biopsy instruments. ENB does not overcome limitations of inadequate specimen collection and processing. Yield may be further increased with more flexible dedicated peripheral instruments. In future applications, ENB may be used with fusion CT/PET for procedure planning; in addition, guided instruments may further extend the utility of this new technology.
Dr. Karnak's participation was supported by the Short-Term Scientist Exchange Program, Office of International Affairs, National Cancer Institute/USA and TUBITAK/TURKEY. This article has an online supplement, which is accessible from this issue's table of contents at www.atsjournals.org Originally Published in Press as DOI: 10.1164/rccm.200603-344OC on July 27, 2006 Conflict of Interest Statement: T.R.G. has received less than $5,000 for serving as an "expert clinician" in the application for this technology to the Center for Medicare Services. P.J.M. does not have a financial relationship with a commercial entity that has an interest in the subject of this manuscript. D.K. does not have a financial relationship with a commercial entity that has an interest in the subject of this manuscript. M.M. has participated in the application for this technology to the Center for Medicare Services as an "Expert" and received $6,000. A.C.M. has stock options in superDimension Ltd, but has received no monetary compensation for his involvement with the company. He will receive compensation if the company becomes a publicly traded entity. Recognizing this potential perceived conflict, Dr. Metha had no influence on the design, conduct, enrollment or performance of procedure. He is a member of the scientific advisory board of superDimension Ltd, but has received no monetary compensation for this activity. Received in original form March 8, 2006; accepted in final form July 25, 2006