A Barrier to the Transport of Nanospheres |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Cystic fibrosis (CF) is characterized by the presence of a viscoelastic mucus layer in the upper airways and bronchi. The underlying problem is a mutation in the gene encoding the cystic fibrosis transmembrane conductance regulator protein. Clinical studies of gene transfer for CF are ongoing. For gene delivery to the airways of CF patients to be effective, the mucus covering the target cells must be overcome. We therefore examined the extent to which CF sputum presents a physical barrier to the transport of nanospheres of a size comparable to that of lipoplexes and other transfection systems currently being clinically evaluated for CF gene therapy. We observed that an extremely low percentage of nanospheres (< 0.3%) moved through a 220-µm-thick CF sputum layer after 150 min. The largest nanospheres studied (560 nm) were almost completely blocked by the sputum, whereas the smaller nanospheres (124 nm) were retarded only by a factor of 1.3 as compared with buffer. Surprisingly, the nanospheres diffused significantly more easily through the more viscoelastic sputum samples. We hypothesize that the structure of the network in sputum becomes more macroporous when the sputum becomes more viscoelastic. Sputum from a patient with chronic obstructive pulmonary disease retarded the transport of nanospheres to the same extent as did CF sputum. When directly mixed with CF sputum, recombinant human deoxyribonuclease I moderately facilitated the transport of nanospheres through CF sputum.
| |
INTRODUCTION |
|---|
|
|
|---|
The underlying problem in cystic fibrosis (CF) is a mutation in
the gene encoding the cystic fibrosis transmembrane conductance regulator (CFTR) protein. This protein, which acts as a
Cl
channel, is localized in exocrine glands and secretory epithelia. A defective CFTR protein in the respiratory tract reduces the secretion of Cl
in the overlying mucus layer and enhances Na+ absorption by epithelial cells (1). This alters the
appearance and composition of respiratory mucus and leads to
the respiratory complications of CF, which include airway obstruction, chronic lung infection, and inflammatory reactions.
Current treatment of CF at the respiratory level is directed at
reducing the seriousness of these complications (1). However,
such treatment does not change the mutation in the CFTR
gene that is the basic defect in CF.
In CF gene therapy, either viral or nonviral transfection vectors are used. Although viral vectors are clearly advantageous with respect to transfection efficiency, they may have some disadvantages, such as the induction of an immune response, cytopathic effects, and the possibility of recombination with wild-type viruses (2). Attention has therefore recently been given to nonviral vectors, and in particular to cationic liposomes that can spontaneously form complexes with nucleic acids (2). This association results in the formation of so-called lipoplexes (5).
A limited number of clinical trials of CF gene therapy have
been reported (6). With regard to lipoplexes, only transfection of the nasal epitheluim has been clinically evaluated (6). However, the outcome of the first clinical trial of lipoplexes, in
which these complexes were nebulized into the lungs of CF
patients, was recently reported (4). All clinical studies of CF
gene therapy have reported only a very limited correction of
Cl
transport. Moreover, in vivo transfection appears to be a
thousand times less efficient than in vitro transfection (7). This
might be attributable to differences in cell characteristics as
well as to the presence of noncellular barriers. Because the
target cells for gene therapy in CF are the submucosal gland
cells and the epithelial cells lining the small conducting airways in the lungs (7, 8), gene transfection systems must permeate the tenacious overlying mucus layer in order to be effective. As compared with normal airway secretions, CF mucus has
a higher consistency because of a high content of actin, serum
proteins, DNA, alginate, and rigidifying lipids (9). These
negatively charged biopolymers (mucin, DNA, and alginate)
are connected to each other through physical entanglements
of their chains and noncovalent interactions such as electrostatic, hydrogen, and hydrophobic bonding. The result is a viscoelastic network. Because it is well known that the mucus
blanket in CF protects the epithelial cells by forming a diffusion barrier for pathogens and harmful particles (12), and because it has been observed that gastrointestinal mucus retards the diffusion of macromolecules (13), we wondered about the extent to which the large (100 to 500 nm), positively charged lipoplexes (14) and other systems used for transfection in CF gene therapy would be able to cross the CF sputum layer. No
information about the diffusion of such large particles through
pathologic respiratory sputum is available. However, it has recently been suggested that cationic liposomes and adenovirus-mediated transfection in vitro are significantly reduced in the
presence of CF sputum (7, 15).
Because we were primarily interested in the influence of particle size and the viscoelasticity of CF sputum on the transport of nanosized particles through such sputum, we used inert fluorescence-labeled polystyrene nanospheres of various sizes as a model for gene particles. Our study was aimed at answering the following questions: (1) To what degree are nanospheres retarded by CF sputum and COPD sputum? (2) Is there a relation between the number of nanospheres that move through CF sputum and the viscoelasticity of the sputum? (3) What is the effect of the size of the nanospheres on their passage through CF sputum? and (4) Does the mucolytic agent recombinant human deoxyribonuclease (rhDNase) I enhance the transport of nanospheres through CF sputum?
| |
METHODS |
|---|
|
|
|---|
Sputum
After approval by the Ethics Committee of the University Hospital of
Ghent, sputum was obtained from CF patients and from a patient
with COPD who attended the hospital. The sputum was produced
spontaneously during respiratory therapy. After collection, sputum
samples were immediately frozen at
20° C. Sputum viscoelasticity was measured with a controlled-stress rheometer (AR 1000-N; TA-Instruments, Ghent, Belgium) at 20° C, using a cone-plate geometry.
The angle between the cone and the plate was 2 degrees, and the sample volume required was approximately 0.9 ml. Dynamic oscillatory
measurements were made at a constant frequency of 1 Hz, with a
stress ranging from 0.01 to 0.1 Pa. To avoid disruption of the weak
biopolymer network in sputum, the elastic (G') and viscous (G
) moduli of the sputum samples were determined in the linear viscoelastic region.
The concentration of DNA in the sputum samples was determined fluorometrically with a modified diaminobenzoic acid (DABA) assay (16). Sputum pretreated with 10 µg rhDNase I (Pulmozyme; Roche, Brussels, Belgium; 1 µg rhDNase I = 1 U) per milliliter of sputum was diluted 20-fold with buffer (0.01 M Na2HPO4, 0.04% NaN3; pH 7.4) and placed in a sonification bath for 15 min for further dispersion. The dispersed sputum was incubated at 60° C for 1 h. Subsequently, 300 µl of diluted sputum was incubated with 300 µl of 20% DABA solution at 60° C. After 1 h, the reaction was stopped by adding 10 ml of 1.76 M HCl. The fluorescence of the reaction product was measured at an excitation wavelength of 390 nm and emission wavelength of 530 nm (SLM-Aminco Bowman; Spectronic Instruments Inc., Rochester, NY).
To determine its mucin concentration, sputum was also pretreated with 10 µg rhDNase I per milliliter of sputum. The sputum was then diluted 200-fold in buffer (0.01 M Na2HPO4, 0.04% NaN3; pH 7.4) and placed in a sonication bath for 15 min for further dispersion. Subsequently, 200 µl of the diluted sputum was incubated with 240 µl of an alkaline solution of 0.6 M 2-cyanoacetamide at 100° C for 30 min. After incubation, 10 ml of 0.06 M borate buffer at pH 8.0 was added. A borate buffer was used because it was observed that the fluorescence of the reaction product was enhanced in the presence of borate ions (17). The fluorescence was measured at an excitation wavelength of 336 nm and emission wavelength of 383 nm.
Fluorescent Nanospheres
Fluorescent (yellow-green) polystyrene nanospheres of different sizes,
bearing carboxyl groups on their surface, were purchased from Molecular Probes (Eugene, OR). The average particle size of the nanospheres was measured by photon correlation spectroscopy (PCS), using
an Autosizer 4700 (Malvern, Worcestershire, UK). For this purpose,
the commercial nanosphere suspension was diluted with phosphate
buffer (0.01 M Na2HPO4, 0.04% NaN3; pH 7.4). The weight-average
hydrodynamic diameters of the nanospheres were 124 ± 2 nm (mean ± SD), 270 ± 6 nm, and 560 ± 11 nm. The zeta potentials of the nanospheres were determined with a Zetasizer 2000 (Malvern, Worcestershire, UK), and were
50 ± 2 mV,
47 ± 4 mV, and
57 ± 1 mV for
the 124-nm, 270-nm, and 560-nm nanospheres, respectively.
Transport of Nanospheres through Sputum
Transport of the fluorescent nanospheres through CF and COPD sputum was studied at 20 ± 1° C with a vertical diffusion chamber system (Corning-Costar Corp., Cambridge, MA). As shown in Figure 1, this equipment was modified to a tricompartimental system as developed by Norris and colleagues (18). The modified system consisted of a donor side and an acceptor side, separated from each other by a layer of CF or COPD sputum (220 µm thick, 1.13 cm2 area). The sputum was held in place with the aid of two track-etched porous polycarbonate membranes (Corning-Costar) with an average pore size of 3 µm. The donor compartment was filled with 5.0 ml of nanosphere dispersion (6.8 × 109 nanospheres/ml in 0.01 M Na2HPO4, 0.04% NaN3, pH 7.4). To prevent a difference in hydraulic pressure across the thin sputum layer upon filling of the donor compartment with nanosphere dispersion, the acceptor compartment was simultaneously filled with 5.0 ml buffer. Mixing was done by a flow of nitrogen through both compartments. The transport of nanospheres through the sputum was determined by measuring the nanosphere concentration at the acceptor side of the system after 150 min. This concentration was determined fluorometrically by taking 1.0 ml of sputum out of the acceptor side of the system. To again prevent a difference in hydraulic pressure across the thin sputum layer, 1.0 ml of sputum was simultaneously pipetted out of the donor side of the system. After measurement of their fluorescence, the 1.0-ml volumes of sputum were returned to the donor and acceptor compartments, respectively, of the system.
|
To validate the transport experiments, we evaluated the binding of nanospheres to the acrylic walls of the diffusion chamber and the size of the nanospheres during the time course of the experiments. We also determined the leakage of mucin and DNA from the sputum layer, and the effect of rhDNase I on this leakage. The loss of mucin and DNA from the sputum layer was measured by simultaneously adding 5.0 ml of buffer to the donor and acceptor compartments. Mixing was again done by a flow of nitrogen through both compartments. After 150 min, the solutions in both compartments were collected. After evaporation of the solutions, the dry material was respectively solubilized in 300 µl of buffer when the DNA concentration was measured, and in 240 µl buffer when the mucin concentration was measured. We also determined the mucin and DNA concentrations of the sputum samples used in the transport experiments, in order to allow us to calculate the percentage of loss.
Influence of rhDNase I on Transport of Nanospheres through CF Sputum
To evaluate the effect of rhDNase I on the transport of the 270-nm
nanospheres through CF sputum, we performed two types of experiments on sputum Samples 5 and 6 in Table 2. In a first series of experiments, we added rhDNase I to the nanosphere dispersion in the
donor compartment of the diffusion chamber system to obtain final
rhDNase I concentrations of 10, 30, and 60 µg/ml, respectively. In a
second series of experiments, we mixed rhDNase I directly with the
sputum before it was placed between the acceptor and donor compartments. The final rhDNase I concentration in the CF sputum was
6 µg/ml, a concentration observed in CF sputum after aerosolization
of rhDNase I in CF patients (19). Upon mixing rhDNase I into the
sputum, we followed mucolysis rheologically until no further mucolysis occurred. This revealed a reduction of G' and G
of 50% and 75%,
respectively, for sputum Samples 5 and 6. Subsequently, the degraded
sputa were placed between the donor and acceptor sides of the diffusion chamber system. The nanosphere concentration in the donor
compartment was again 6.8 × 109 spheres/ml, and transport was followed for 150 min.
|
Confocal Microscopy
Confocal scanning laser microscopy (CSLM) (MRC1024; Bio-Rad, Hemel Hempstead, UK) was used to confirm the increased transport of nanospheres through the rhDNase I-pretreated CF sputa. After diffusion of the nanospheres, the 220-µm-thick sputum layers (Figure 1) were removed from the diffusion chamber system. The polycarbonate membranes were carefully washed with distilled water. Consequently, the Snapwell holding the sputum layer was placed on a cover slip. Confocal images (154 × 154 µm) of the nanospheres in the sputa were made at 10 µm below the surface of the sputum layer that had been in contact with the nanosphere dispersion in the donor compartment during the transport experiments. The number of fluorescent nanospheres in a microregion of each sputum sample was counted with appropriate image analysis software.
Electron Microscopy
CF sputum samples were processed for scanning electron microscopic examination according to a protocol used for the study of uterine cervical mucus (20). Samples were fixed for 1 h in 2% glutaraldehyde in phosphate-buffered saline (PBS) (2.5 mM KH2PO4, 7.0 mM K2HPO4, 0.12 M NaCl, pH 7.6), subsequently rinsed in buffer, and postfixed for 30 min in 1% OsO4 in PBS. After three washes in PBS, the sputum samples were dehydrated through a series of ethanol solutions, critical-point dried with liquid CO2, and sputter-coated with platinum. These prepared CF sputum samples were then examined with a scanning electron microscope (JSM-5000 LV; JEOL, Tokyo, Japan).
Statistical Analysis
The experimental results in this report are expressed as mean ± SD. Spearman's rank correlation test was used to study the relation between sputum viscoelasticity and sputum DNA and mucin content. In studying the effect of sputum viscoelasticity on nanosphere transport, we tested the equality of means through analysis of variance. Student's t test was used in evaluating the effect of rhDNase I on nanosphere transport. The t test was also used to evaluate the rheologic stability of CF sputum as a function of temperature, and to evaluate the effect of freezing on sputum viscoelasticity. Significance was set at p = 0.05.
| |
RESULTS |
|---|
|
|
|---|
Because the sputum samples were stored at
20° C, we evaluated the effect of freezing and thawing on their viscoelasticity. Table 1 shows the rheologic properties of eight sputum samples, all from different CF patients, before and after freezing
at
20° C in airtight containers for 3 wk. No significant change
in G' or G
could be detected (p > 0.05).
|
Since we were interested in the influence of intact, nondegraded sputum on the transport of nanospheres, and because
the transport experiments took about 150 min, we evaluated
the rheologic stability of CF sputa and the COPD sputum
sample. Sputum may undergo physical degradation after its
collection, through the action of endogenous and bacterial
proteases (21). This enzymatic degradation, which results in
liquefaction of the sputum by cleavage of the protein backbone of mucin biopolymers, is often ignored in drug transport studies through mucus or sputum. To evaluate the rheologic
stability of the CF sputum samples, we placed them between
the cone and plate of the rheometer and made measurements
of G' and G
at 20° C and 37° C, respectively. To avoid sputum
dehydratation from water evaporation, we used a solvent trap.
The sputum viscoelasticity dropped by more than 30% after 2 h
at 37° C, even in the presence of 0.21 mg/ml phenylsulfonylfluoride, a protease inhibitor, and 0.40 mg/ml NaN3, an antibacterial agent. In contrast, no significant change in G' and G
was
observed after 3 h when CF sputum samples were kept at 20° C
in the presence of phenylsulfonylfluoride (0.21 mg/ml) and
NaN3 (0.40 mg/ml) (p > 0.05). The COPD sputum sample also
showed no change in viscoelasticity under these conditions. On the basis of these results, we conducted the transport experiments at 20° C for 150 min while carefully adding phenylsulfonylfluoride (0.21 mg/ml) and NaN3 (0.40 mg/ml) to the
sputum samples.
Table 2 shows the rheologic moduli of the six CF sputum
samples, all from different patients, and of the COPD sputum
sample used in the transport experiments. In the case of the
CF sputum samples, each value is based on the rheologic analysis of three sputum fractions of each sample. Because of the
limited amount of COPD sputum, only one fraction could be
taken of this sample. Notwithstanding the visually observed
heterogeneity of the CF sputum, a good reproducibility of G'
and G
was obtained (Table 2). This may be attributed to the
large volume fractions (0.9 ml) taken from the CF sputum samples used in the determination of G' and G
. In this way, the
heterogeneity of the samples became more averaged. Because
G' of each sputum sample is significantly larger than G
, CF
sputum can be considered as a dominantly elastic biomaterial.
The DNA and mucin contents of the CF sputa and the
COPD sputum sample are shown in Table 2. A significant correlation was found between mucin content and the rheologic
moduli G' and G
(rs = 0.83; p = 0.04). Additionally, a positive correlation, although not significant, was observed between the DNA concentration and the viscoelasticity of the
CF sputa (rs = 0.37; p = 0.47). Because a modified DABA assay was used, only the total DNA content of the sputum samples was measured. These data gave no information about the
length of the DNA chains in CF sputum, which may play a major role in increasing its viscoelasticity. This might explain the
lack of a significant correlation between the DNA content and
viscoelasticity of the CF sputa.
During the transport study, no significant change in particle
size was observed for any type of nanospheres used. The loss of mucin and DNA from the sputa into the donor and acceptor compartments of the diffusion chamber system depended
on the addition of rhDNase I to the sputa. The loss of mucin
from a first sputum sample (G' = 22 ± 6 Pa; G
= 6 ± 2 Pa)
increased from 7 ± 1% (before treatment with rhDNase I) to
15 ± 2% upon adding rhDNase I (6 µg/ml). For a second CF
sputum sample (G' = 10 ± 3 Pa; G
= 3 ± 1 Pa) the loss of
mucin increased from 4 ± 2% to 13 ± 5%. To determine the loss of DNA, two other sputum samples were used. The leakage of DNA from the first sputum sample (G' = 17 ± 6 Pa;
G
= 4 ± 1 Pa) increased from 5 ± 1% (before treatment with
rhDNase I) to 16 ± 5% upon adding rhDNase I (6 µg/ml). For
the second CF sputum sample (G' = 13 ± 3 Pa and 3 ± 1 Pa)
the leakage increased from 7 ± 1% to 14 ± 2%. Therefore, in
the absence of rhDNase I, only a small amount of mucin and
DNA leaked into the chambers. However, mixing rhDNase I
with the sputa significantly increased these amounts of mucin
and DNA. This was not surprising, knowing that rhDNase I
cleaves DNA chains and thereby causes a weakening of the
sputum viscoelastic network.
The binding experiments showed that the smallest nanospheres (124 nm) clung to the acrylic wall of the acceptor compartment of the diffusion chamber system, whereas the 270-nm and 560-nm nanospheres did not cling. To correct the concentration of the 124-nm nanospheres in the acceptor compartment for nonspecific adsorption, we performed additional experiments. The acceptor compartment was filled with 124-nm nanosphere dispersions, which, after adsorption, resulted in approximately the same nanosphere concentrations as measured in the transport experiments on the acceptor side of the diffusion chamber system after 150 min. From this decrease in nanosphere concentration we calculated, at each measuring time, a correction factor. To obtain the real quantity of 124-nm nanospheres that had moved through the sputum layer, we multiplied the experimentally determined 124-nm nanosphere concentrations in the acceptor compartment by the corresponding correction factors.
Transport experiments with the nanospheres were initially done on sputum Samples 1 through 4 in Table 2. Additional transport experiments were done with the 270-nm nanospheres on sputum Samples 5 and 6 and on the COPD sample. Each transport experiment was repeated four times. Figure 2 shows the percentages of nanospheres that were transported into the acceptor compartment after 150 min for the 124-nm, 270-nm, and 560-nm nanospheres, respectively. For each sputum sample, we observed a nonlinear decrease in the percentage of transported nanospheres as a function of the size of the nanospheres. The mean percents of nanospheres transported after 150 min through the four CF sputum samples were 0.24 ± 0.08% (124 nm), 0.022 ± 0.008% (270 nm), and 0.0017 ± 0.0009% (560 nm), respectively. On the basis of the size of the nanospheres, we calculated the percentage of nanospheres that could be expected in the acceptor compartment after 150 min with buffer present instead of a sputum layer between the donor and acceptor compartments (22). The calculated percents were 0.32%, 0.15%, and 0.071% for the 124-nm, 270-nm, and 560-nm nanospheres, respectively.
|
Figure 3 shows the percents of nanospheres that were
transported into the acceptor compartment of the diffusion
chamber system after 150 min as a function of G' of the CF
sputa. Similar profiles were obtained when the G
was plotted
on the x axis. Surprisingly, for the 124-nm and 270-nm nanospheres, the percentages of transported nanospheres through
the most tenacious sputum sample, Sample 4, was significantly
greater than the percentages of nanospheres transported
through the less viscoelastic sputum samples (p < 0.05). Additional transport studies (using the 270-nm nanospheres) with
CF sputum Samples 5 and 6 (Table 2) confirmed that there was less nanosphere transport through less viscoelastic CF
sputum samples (Figures 2 and 3). This phenomenon was also
observed for the COPD sputum sample.
|
The COPD sputum sample showed a much lower DNA concentration and viscoelasticity than did the CF sputa, whereas the mucin content of the two kinds of sputum samples was similar (Table 2). This confirms that a main reason for the high viscoelasticity of CF sputum is the presence of massive amounts of DNA. To reduce sputum viscoelasticity, rhDNase I is frequently used in treating CF. Because rhDNase I partly disrupts the biopolymer network in CF sputum, we wondered whether it would enhance the transport of nanospheres. Figure 4 shows the results. When rhDNase I was present on the donor side of the diffusion chamber system during the transport experiments, the greatest increase in transport was observed when a rhDNase I concentration of 60 µg/ml was used. When using lower rhDNase I concentrations (10 µg/ml and 30 µg/ml) on the donor side of the system, the transport of nanospheres was only moderately enhanced. The strongest increase in transport was seen when rhDNase I was mixed directly with CF sputa, even at a low concentration of 6 µg/ml. The increase in nanosphere transport upon mixing rhDNase I with the sputa was confirmed by confocal scanning laser microscopy of the CF sputum layers that were removed from the diffusion chamber system at the end of the transport experiments. Figures 5A and 5B show representative microregions of a CF sputum layer in which rhDNase I was mixed with the sputum before the transport experiment and when it was absent, respectively. The microregion of the sputum that had been mixed with rhDNase I contained 95 nanospheres, whereas only 14 nanospheres were present in an equal-size microregion of sputum with which no rhDNase I was mixed.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
It was our goal to study the effect of particle size, sputum viscoelasticity, and rhDNase I on the transport of nanosized particles through CF and COPD sputum. In this way we hoped to be able to more fully elucidate the extent to which sputum represents a physical barrier to the transport of nanosized particles. Moreover, we hoped to provide an additional viewpoint to that of recently reported studies (7, 15) of the effect of CF sputum on in vitro gene transfection efficiency by lipoplexes and adenoviral carriers.
In our study, we used negatively charged polystyrene nanospheres because it is known that in vivo, cationic gene complexes are less efficient than anionic or neutral gene complexes (23). This lower efficacy of positively charged gene complexes is attributed to electrostatic interactions with anionic biopolymers in vivo (24).
CF sputa dramatically retarded the transport of large particles (Figures 2 and 3). We showed that the influence of surface hydrophobicity on the transport of nanospheres was much smaller than the effect of particle size (data not shown). We therefore concluded that the difference in transport of 124-nm, 270-nm, and 560-nm nanospheres was mainly caused by stronger steric obstruction with increasing particle size.
As compared with diffusion through buffer, the quantities of nanospheres that diffused through CF sputum were 1.3-fold, 6.8-fold and 42-fold smaller for the 124-nm, 270-nm and 560-nm nanospheres, respectively. Because the smallest nanospheres were retarded only by a factor of 1.3, it appears that the water channels in CF sputum were large enough to allow passage of these spheres. This indicates that the limited transport of the smallest nanospheres in CF sputum was mainly due to the distance they had to travel before they reached the acceptor compartment of the diffusion chamber system used in the study. On the other hand, the 560-nm nanospheres were almost completely blocked sterically by the CF sputum. This agreed with our electron microscopic data for CF sputa (Figure 6), which revealed pores with a diameter ranging from approximately 100 nm to 400 nm. However, it should be noted that distortions may have occurred in the network structure of the sputum during the fixation and dehydration steps of the sample preparation for electron microscopy.
|
Generally, diffusion slows when gels become more viscoelastic (25). This contrasts with the findings of the present study. As shown in Figure 3, a greater number of nanospheres moved through the more viscoelastic sputa. To explain these results, it is necessary to consider the structure of the three- dimensional network in sputum. For cervical mucus it has been proposed that the mucin chains, owing to their ordered arrangement, form a three-dimensional, macroporous network of channels whose large pores permit macromolecular drug transport (20). It has been mentioned that liposomes (~ 200 nm) can move through gastrointestinal mucus (13), whereas the transport of particles larger than 500 nm is almost completely blocked (18, 26). Additionally, Yudin and coworkers showed that in cervical mucus, a fine structure with interfiber spacings of ~ 100 nm is present within a more macroporous structure with interfiber spacings of ~ 500 nm (27). On the basis of this information, the following hypothesis may explain the greater transport of nanospheres through more viscoelastic sputa: An increase in the viscoelasticity of sputum originates from a high number of junctions between biopolymer chains per volume unit of sputum. An increase in the concentration of junctions may originate from an increase in the concentration of biopolymers in the sputum. Indeed, Table 2 shows a significant positive correlation between G' and mucin concentration (rs = 0.83; p = 0.04) and a positive correlation between G' and DNA concentration (rs = 0.37; p = 0.47). For synthetic gels it was observed that increasing the concentration of junctions may change the network structure of the gel from one of homogeneous microporous type to one of more heterogeneous macroporous type (28). A similar phenomenon may occur in biologic gels such as sputum. In weakly viscoelastic CF and COPD sputum, a more homogeneous microporous network may be present, with many free biopolymer chains in the aqueous channels of the network. By contrast, a more heterogeneous macroporous network may be established in highly viscoelastic sputum, with a limited number of free biopolymer chains in its aqueous channels. Both the macroporous channels and the low number of free biopolymer chains may facilitate the transport of nanospheres in the more viscoelastic sputum.
The experiments on the transport of 270-nm spheres through the COPD sputum sample revealed that the transport characteristics, as shown in Figures 2 and 3, are not unique to CF sputum. In particular, the numbers of nanospheres transported through the COPD sputum were fully in accord with the influence of viscoelasticity (Figures 2 and 3).
Figure 4 shows that when rhDNase I was added to the donor compartment of the diffusion chamber system used in the study, only a small increase occurred in nanosphere transport through CF sputum. The greatest, albeit only a limited, increase in nanosphere transport occurred when rhDNase I was directly mixed with CF sputum (6 µg/ml) before the sputum was placed between the donor and acceptor compartments of the diffusion chamber system. The limited enhancement of transport may be explained by the presence of mucins and other biopolymers as well as DNA. Consequently, rhDNase I does not completely break down the biopolymer network of sputum. Moreover, the smaller DNA fragments that result from DNA degradation may themselves hinder the transport of nanospheres through viscous effects. The loss of DNA and mucin from the sputum layer, which was significantly greater when rhDNase I was mixed with the sputum, may have enhanced the transport.
Assuming that the penetration of rhDNase I through CF sputum was rapid when a concentration of 10 µg/ml rhDNase I was added to the donor compartment, it can be expected that equilibrium of rhDNase I in such sputum will occur at a concentration of approximately 5 µg/ml sputum. This would suggest a similar effect of rhDNase I on nanosphere transport to the effect observed when 6 µg of rhDNase I was directly mixed with 1.0 ml sputum. Figure 4 shows that this was not the case. Because no loss of rhDNase I activity in the donor compartment of the diffusion chamber system was observed during the transport experiment, denaturation of rhDNase cannot explain the weak effect on nanosphere transport. Weak penetration of rhDNase I from the donor side of the system into the sputum layer may explain the weak effect on nanosphere transport. However, another explanation is that rhDNase I may have been inactive in the sputum because calcium and magnesium ions, which are necessary for optimal activity of rhDNase I (29), diffused from the sputum into the donor and acceptor compartments of the diffusion chamber system during the transport experiments. Because the addition of calcium and magnesium ions to the nanosphere dispersion might have destabilized the dispersion, we had to exclude these ions from our buffers.
Given that rhDNase I decreases sputum viscoelasticity, it can be asked, on the basis of the data shown in Figure 3, why a decrease in nanosphere transport was not observed in the presence of rhDNase I. This may be explained by the dualistic network structure of sputum as discussed earlier. In the macroporous network, the DNA chains may participate in ordered bundles of polymers that may be less accessible to degradation by rhDNase I than are DNA chains that are present in the homogeneous microporous network. Consequently, rhDNase I may first degrade the more accessible DNA chains in the microporous network, which may reduce the viscous drag on the diffusion of nanospheres, and may explain their increased transport through CF sputum.
In this study, we showed that CF sputum offers a size-dependent barrier to the transport of nanospheres of a size comparable to that of lipoplexes and other transfection systems. As compared with their diffusion through buffer, the smallest nanospheres (124 nm) used in our study were only moderately retarded by CF sputum. However, the larger, 560-nm nanospheres were almost completely blocked by CF sputum. This difference in transport was caused mainly by greater steric obstruction of the largest particles in the sputum. Our study also demonstrated that nanospheres moved more easily through sputum samples with a greater viscoelasticity. To explain this unexpected observation, we hypothesized that the network in CF sputum changes toward having a more macroporous structure when the sputum becomes more viscoelastic. COPD sputum retarded the transport of nanospheres to the same extent as did CF sputum. The strongest enhancement of the transport of nanospheres through CF sputum by rhDNase I, albeit by only a factor of 3, was observed when this mucolytic agent was mixed with the sputum. When rhDNase I was presented on the surface of CF sputum, as occurs in vivo, only a very small increase was observed in nanosphere transport through the sputum.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Stefaan De Smedt, Laboratory of General Biochemistry and Physical Pharmacy, Faculty of Pharmacy, Ghent University, Harelbekestraat 72, 9000 Ghent, Belgium. E-mail: Stefaan.Desmedt{at}rug.ac.be
(Received in original form September 9, 1999 and in revised form March 15, 2000).
Niek Sanders is a doctoral fellow of Vlaams Institut voor de bevordering vat het Wetenschappelijk-Technologisch Onderzoek in de Industrie. The financial support of this institute is acknowledged with gratitude.Acknowledgments: The vertical diffusion chamber system used in the study was a gift from Corning-Costar, Inc., Cambridge, MA. The authors thank Ghent University and Fonds voor Wetenschappelijk Onderzoek-Vlaanderen for their support through instrumentation credits (Autosizer 4700, Malvern; Zetasizer 2000, Malvern; AR 1000 N, TA Instruments; Aminco Bowman, Spectronic Instruments Inc.; FWO/KNV 315.563.98). They thank Prof. Dr. K. De Boeck of the Catholic University of Leuven, Belgium, for providing them with CF sputum samples. They also thank Prof. Dr. P. Van Oostveldt (Ghent University, Belgium) for the use of the MRC1024 Bio-Rad CSLM.
| |
References |
|---|
|
|
|---|
1. Davis PB, Drumm M, Konstan MW. Cystic fibrosis. Am J Respir Crit Care Med 1996; 154: 1229-1256 [Medline].
2. Lee RJ, Huang L. Lipidic vector systems for gene transfer. Crit Rev Ther Drug Carrier Syst 1997; 14: 173-206 [Medline].
3. Chadwick SL, Kingston HD, Stern M, Cook RM, OConnor BJ, Lukasson M, Balfour RP, Rosenberg M, Cheng SJ, Smith AE, et al . Safety of a single aerosol administration of escalating doses of the cationic lipid GL-67/DOPE/DMPE-PEG5000 formulation to the lungs of normal volunteers. Gene Ther 1997; 4: 937-942 [Medline].
4. Alton EWFW, Stern M, Farley R, Jaffe A, Chadwick SL, Phillips J, Davies J, Smith SN, Browning J, Davies MG. et al. Cationic lipid-mediated CFTR gene transfer to the lungs and nose of patients with cystic fibrosis: a double-blind placebo-controlled trial. Lancet 1999; 353: 947-954 [Medline].
5. Felgner PL, Barenholz Y, Behr JP, Cheng SH, Cullis P, Huang L, Jessee JA, Seymour L, Szoka F, Thierry AR, et al . Nomenclature for synthetic gene delivery systems. Hum Gene Ther 1997; 8: 511-512 [Medline].
6. Geddes D, Alton E. Cystic fibrosis clinical trials. Advanced Drug Delivery Rev 1998; 30: 205-217 . [Medline]
7. Kitson C, Angel B, Judd D, Rothery S, Severs NJ, Dewar A, Huang L, Wadsworth SC, Cheng SH, Geddes DM, et al . The extra- and intracellular barriers to lipid and adenovirus-mediated pulmonary gene transfer in native sheep airway epithelium. Gene Ther 1999; 6: 534-546 [Medline].
8. Boucher RC. Status of gene therapy for cystic fibrosis lung disease. J Clin Invest 1999; 103: 441-445 [Medline].
9. Sheils CA, Käs J, Travassos W, Allen PG, Janmey PA, Wohl ME, Stossel TP. Actin filaments mediate DNA fiber formation in chronic inflammatory airway disease. Am J Pathol 1996; 148: 919-927 [Abstract].
10. Mrsny RJ, Daugherty AL, Short SM, Widmer R, Siegel MW, Keller GA. Distribution of DNA and alginate in purulent cystic fibrosis sputum: implications to pulmonary targeting strategies. J Drug Target 1996; 4: 233-243 [Medline].
11. Girod S, Galabert C, Lecuire A, Zahm JM, Puchelle E. Phospholipid composition and surface-active properties of tracheobronchial secretions from patients with cystic fibrosis and chronic obstructive pulmonary diseases. Pediatr Pulmonol 1992; 13: 22-27 [Medline].
12. Girod S, Zahm J-M, Plotkowski C, Beck G, Puchelle E. Role of the physicochemical properties of mucus in the protection of the respiratory epithelium. Eur Respir J 1992; 5: 477-487 [Abstract].
13. Desai MA, Mutlu M, Vadgama P. A study of macromolecular diffusion through native porcine mucus. Experientia 1992; 48: 22-26 [Medline].
14. Eastman SJ, Siegel C, Tousignant J, Smith AE, Cheng SH, Scheule RK. Biophysical characterization of cationic lipid:DNA complexes. Biochim Biophys Biomembr 1997; 1325: 41-62 .
15. Stern M, Caplen NJ, Browning JE, Griesenbach U, Sorgi F, Huang L, Gruenert DC, Marriot C, Crystal RG, Geddes DM, et al . The effect of mucolytic agents on gene transfer across a CF sputum barrier in vitro. Gene Ther 1998; 5: 91-98 [Medline].
16. Puchelle E, Zahm J-M, de Bentzmann S, Grosskopf C, Shak S, Mougel D, Polu J-M. Effects of rhDNase on purulent airways secretions in chronic bronchitis. Eur Respir J 1996; 9: 765-769 [Abstract].
17. Miquel JF, Groen AK, van Wijland MJA, del Pozo R, Eder MI, von-Ritter C. Quantification of mucin in human gallbladder bile: a fast, specific, and reproducible method. J Lipid Res 1995; 36: 2450-2458 [Abstract].
18. Norris DA, Sinko PJ. Effect of size, surface charge, and hydrophobicity on the translocation of polystyrene microspheres through gastrointestinal mucin. J Appl Polymer Sci 1997; 63: 1481-1492 .
19. Sinicropi MW, Prince WS, Lofgren JA, Lucas M, Devault A, Baughman S, Nash M, Fuchs H, Shak S. Sputum pharmacodynamics and pharmacokinetics of recombinant DNase I in cystic fibrosis [abstract]. Am J Respir Crit Care Med 1994; 149: A671 .
20. Saltzman WM, Radomsky ML, Whaley KJ, Cone RA. Antibody diffusion in human cervical mucus. Biophys J 1994; 66: 508-515 [Medline].
21. Jacquot J, Hayem A, Galabert C. Functions of proteins and lipids in airway secretions. Eur Respir J 1992; 5: 343-358 [Abstract].
22. Crank J. The mathematics of diffusion, 2nd ed. Oxford: Clarendom Press; 1975. p. 44-68.
23. Ferrari S, Moro E, Pettenazzo A, Behr JP, Zacchello F, Scarpa M. ExGen 500 is an efficient vector for gene delivery to lung epithelial cells in vitro and in vivo. Gene Ther 1997; 4: 1100-1106 [Medline].
24. Schwartz B, Benoist C, Abdallah B, Scherman D, Behr JP, Demeneix BA. 1995. Lipospermine-based gene transfer into the newborn mouse brain is optimized by a low lipospermine/DNA charge ratio. Hum Gene Ther 1995; 6: 1515-1524 [Medline].
25. De Smedt SC, Meyvis T, Van Oostveldt P, Blonk J, Hennink WE, Demeester J. The diffusion of macromolecules in dextran methacrylate solutions and gels as studied by confocal scanning laser microscopy. Macromolecules 1997; 30: 4863-4870 .
26. Szentkuti L. Light microscopical observations on luminally administered dyes, dextrans, nanospheres and microspheres in the pre-epithelial mucus gel layer of the rat distal colon. J Controlled Release 1997; 46: 233-242 .
27. Yudin A, Hanson F, Katz D. Human cervical mucus and its interaction with sperm: a fine-structural view. Biol Reprod 1989; 40: 661-671 [Abstract].
28. Mallam S, Horkay F, Hecht A, Geissler E. Scattering and swelling properties of inhomogeneous polyacrylamide gels. Macromolecules 1989; 22: 3356-3361 .
29. De Meester I, De Smedt SC. Deoxyribonuclease I. In: Lauwers A, Scharpé S, editors. Pharmaceutical enzymes. New York: Marcel Dekker; 1997. p. 285-304.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |