|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
Biologically variable mechanical ventilation (
bv)
using a computer-controller to mimic the normal variability in spontaneous breathing
improves gas exchange in a model of severe lung
injury (Lefevre, G. R., S. E. Kowalski, L. G. Girling, D. B. Thiessen, W. A. C. Mutch. Am. J. Respir. Crit. Care Med. 1996;154:1567-1572). Improved oxygenation with
bv, in the face of alveolar collapse, is
thought to be due to net volume recruitment secondary to the variability or increased noise in the peak inspiratory airway pressures (Ppaw). Biologically variable noise can be modeled as an inverse power law frequency distribution (y
1/fa) (West, B. J., M. Shlesinger. Am. Sci. 1990;78:40-45). In a porcine model of
atelectasis
right lung collapse with one-lung ventilation
we studied if
bv (n = 7) better reinflates the collapsed lung compared with conventional monotonously regular control mode ventilation (
c; n = 7) over a 5-h period. We also investigated the influence of sigh breaths with
c (
s; n = 8) with this model.
Reinflation of the collapsed lung was significantly enhanced with
bv
greater PaO2 (502 ± 40 mm Hg with
bv versus 381 ± 40 mm Hg with
c at 5 h; and 309 ± 79 mm Hg with
s; mean ± SD),
lower PaCO2 (35 ± 4 mm Hg versus 48 ± 8 mm Hg and 50 ± 8 mm
Hg), lower shunt fraction (9.7 ± 2.7% versus 14.6 ± 2.0% and
22.9 ± 6.0%), and higher respiratory system compliance (Crs)
(1.15 ± 0.15 ml/cm H2O/kg versus 0.79 ± 0.19 ml/cm H2O/kg and
0.77 ± 0.13 ml/cm H2O/kg)
at lower mean Ppaw (15.7 ± 1.4 cm
H2O versus 18.8 ± 2.3 cm H2O and 18.9 ± 2.8 cm H2O).
bv resulted in an 11% increase in measured tidal volume (VTm) over
that seen with
c by 5 h (14.7 ± 1.2 ml/kg versus 13.2 ml/kg). The
respiratory rate variability programmed for
bv demonstrated an
inverse power law frequency distribution ( y
1/fa) with a = 1.6 ± 0.3. These findings provide strong support for the theoretical
model of noisy end-inspiratory pressure better recruiting atelectatic lung. Our results suggest that using natural biologically variable noise has enhanced the performance of a mechanical ventilator in control mode.
| |
INTRODUCTION |
|---|
|
|
|---|
Recruitment of atelectatic lung units and maintenance of alveolar patency is an integral goal of mechanical ventilation. We have recently developed a new mode of mechanical ventilation called biologically variable ventilation (
bv). This computer-controlled ventilator mimics the normal spectrum of
breathing by incorporating breath-to-breath variability in respiratory rate (f) and tidal volume (VT). Using
bv we have
demonstrated improved arterial oxygenation (PaO2) without
an increase in mean airway pressure (
) in a porcine model
of severe lung injury (1). Suki and colleagues (2) postulate
that
bv improves PaO2 owing to recruitment of collapsed alveoli, which open in bursts or avalanches (3). They speculate
that
bv is an example of stochastic resonance
the addition
of noise to an input signal (variable peak airway pressure [Ppaw]) to amplify output (PaO2) in a nonlinear system (4). With the noisy input signal seen with
bv, the volume gained at higher Ppaw greatly exceeded the volume lost at lower pressures over time, the net result being improved oxygenation
without an increase in airway pressure (
).
To test the hypothesis that
bv enhances recruitment of
collapsed alveoli, we developed a porcine model of stable unilateral lung collapse and compared lung reinflation over 5 h
using
bv versus conventional control mode ventilation (
c)
at similar minute ventilation.
We also examined if the addition of sigh breaths to
c was
effective in recruiting collapsed alveoli. We programmed the
sigh breaths to occur at the same interval with the same delivered volume as the largest breaths with
bv. If sighs programmed in this manner were as efficacious as
bv, then the
more complicated variability programmed with
bv would be unnecessary.
| |
METHODS |
|---|
|
|
|---|
The Committee for Animal Experimentation at the University of
Manitoba approved the study. When depth of anesthesia was adequate (isoflurane 1.5 minimal alveolar concentration [MAC] in 100%
O2), a tracheostomy was done and a double-lumen endotracheal tube
was placed in the airway. Correct positioning was confirmed by fibreoptic bronchoscope. Mechanical ventilation by
c was instituted with an Ohio 7000 anesthesia ventilator (Ohio Instruments, Madison, WI)
with f approximately 15 breaths/min and minute ventilation adjusted
to maintain the end-tidal CO2 at approximately 35 mm Hg. Catheters
were placed for blood sampling and pressure measurements. Airway
pressures and volumes were measured by pneumotachograph (Hans
Rudolph, Kansas City, MO). After baseline measurements, the right
side of the double-lumen endotracheal tube was opened to air to allow the right lung to collapse. A minithoracotomy (pleural opening
2.5 cm) permitted complete collapse and observation of the lung. The
lung remained collapsed for 1 h. At this point, the double-lumen tube
was removed and replaced with a cuffed tracheostomy tube for reexpansion of the right lung. Animals were randomly allocated to continue with
c or switched to
bv. Conceptually, this is equivalent to
flipping a switch: on = biological variability; off = no variability. Before the switch was flipped in those animals receiving
bv, f and measured tidal volume (VTm) were the same in each group. The delivered
minute ventilation remained unchanged from baseline and continued
with either mode for the next 5 h. Blood gases and O2 contents (arterial and mixed venous) and expired gas samples were measured (Radiometer ABL3 and Radiometer OSM3, Copenhagen NV, Denmark). Static respiratory system compliance (Crs) was measured by
transiently clamping the expiratory limb of the ventilatory circuit at
end inspiration. Calculated indices included shunt fraction (
S/
T),
and Crs (
V/
P).
Computer-controlled Ventilation
The computer-controller and software for the ventilator have been
previously described (1). Data for the modulation file were obtained
from an awake, spontaneously breathing animal. The variability file
used with
bv is shown in Figure 1.
|
Computer-controlled Sigh Ventilation
Examination of Figure 1 reveals four instantaneous breaths below 10 breaths/min. To deliver computer-controlled sighs, a modulation file
was written so that all instantaneous breaths were set to 15 breaths/
min except at the four time periods where the instantaneous breaths
were < 10 breaths/min. At these times the instantaneous breath rate
was programmed as shown in Figure 1. Thus, each of these low breath
rates resulted in delivery of a large VT or sigh breath at the same interval and magnitude as those programmed for
bv. In addition, the
duration of the modulation file was the same as with
bv before it
looped to repeat itself. Experiments were done post hoc in this group.
Every attempt was made to ensure that this group did not differ at
baseline or with one-lung ventilation from the other two groups.
Post hoc Analysis
Data acquisition files of airway pressure and flows were processed to
integrate the area under the pressure-time and flow-time curves to
give
and volume. Mean Ppaw was also calculated.
Statistical Analysis
Data were analyzed by repeated measures analysis of variance
(ANOVA). A p value
0.05 was considered significant for group × time interactions or differences between groups. Comparisons between and within groups were based on generated least-squares means matrices with Bonferroni's correction applied when multiple comparisons were made. Data are presented as mean ± SD unless otherwise noted.
Inverse power law analysis was done as follows: mean instantaneous f was determined, then each instantaneous f was subtracted from mean f, this value was squared, then log transformed. These data were partitioned into incremental bins of equal size to determine their frequency distribution. The probability of each frequency was determined by Ni/N where Ni = number of observations in a given frequency bin and N = total number of observations. A log transform of the probability distribution was derived. The log probability distribution versus log f variation was plotted. The confidence interval and correlation coefficient were derived by regression analysis.
| |
RESULTS |
|---|
|
|
|---|
Data were analyzed on 22 experiments (n = 7 with
bv and
c and n = 8 with
s). Measured mean f was the same in all
groups at baseline values and during the period of one-lung
ventilation. At these times, there were only minor differences
seen for any measured parameter between groups (Table 1).
Importantly, at baseline there were no differences in PaO2,
PaCO2,
S/
T, Ppaw, VT or Crs between groups. With one-lung ventilation, the PaO2 decreased significantly with approximately a 4-fold increase in
S/
T. The Ppaw nearly doubled.
The Crs decreased to approximately 40% of baseline values.
After 60 min both lungs were again ventilated. In the group
receiving
bv and
s, the computer-controller was activated.
In these two groups, the delivered minute ventilation was not
changed from its baseline settings with
c. Mean f was scaled
to 15 breaths/min, the same mean rate as in the control group
measured rate with
bv remained unchanged from
baseline at 13.9 breaths/min; coefficient of variation 18%. At
5 h, PaO2 was significantly higher with
bv than in the other
two groups [group × time interaction (G × T); p < 0.0001]).
Carbon dioxide clearance was superior with
bv such that
PaCO2 was significantly lower at 5 h than for the other two
groups (G × T; p < 0.0001). Mean Ppaw was lower at 5 h with
bv than with
c or
s (G × T; p < 0.0001). VT was significantly greater with
bv at 5 h (G × T; p < 0.0001). Crs was
much greater with
bv by 5 h than in the other two groups (G × T; p < 0.0001).
|
Figure 2 shows the changes in PaO2 over time for each
group from Time 0 (end one-lung ventilation) to experiment
completion at 5 h. The PaO2 increases more rapidly with
bv
and reaches a higher asymptote (p < 0.0001 group × time interaction) than in the other two groups.
|
At baseline, during
c, minute ventilation, the product of f
and VTm (f × VTm) was not different for
bv versus
c (186 ± 11 ml/kg with
bv and 187 ± 10 ml/kg with
c). With one-lung anesthesia with both groups on
c, measured VT decreased as airway pressure increased. This is compatible with
the volume lost because of the compliance of the anesthesia
circuit (circuit compression volume; 2 ml/cm H2O/L in this
case). With switch to
bv, the measured minute ventilation
product increased despite unchanged ventilator settings. The
f × VTm product increased solely as a consequence of an increase in measured VT. Measured f remains unchanged (> 75 instantaneous breath intervals and VT measured per observation with
bv). Measured VT was approximately 11% greater
with
bv at 5 h compared with
c, such that V
bv = V
c +
V
c with
= 0.11.
In Figure 3 we have plotted log probability distribution versus log variability in f. An inverse power law frequency distribution was seen with slope
1.6 ± 0.3.
|
| |
DISCUSSION |
|---|
|
|
|---|
In this simple experimental model of reversible atelectasis
deflation of one lung in the pig
bv resulted in more rapid and greater recruitment of collapsed lung. Our results are compatible with the theoretical model that "noisy" Ppaw would
better recruit atelectatic lung units (2). With
bv, effective VT
increased 11% under these experimental conditions. Suki and
coworkers showed the probability functions for alveolar recruitment, which occur in avalanches (3, 5). These functions
follow inverse power law frequency distributions (y
1/fa; examples of noise in natural phenomena) (6), with slopes of
1.1 to
2.5. We have plotted the probability distribution of
the variability in f used with
bv (Figure 3). This function also
follows an inverse power law frequency distribution with a
negative slope of 1.6.
Suki and coworkers suggest that "both the magnitude and
timing of pressure excursion applied at the airway entrance
during artificial ventilation may be critical in triggering the avalanche process of alveolar recruitment" (3). As such, variable
f and VT with
bv presumably facilitates this avalanche process. The increase in the f × VTm product is the fundamental
change with
bv. The increase in measured VT is compatible
with the improved Crs seen at 5 h with
bv. Gunnarsson and
coworkers have shown a positive correlation between shunt
fraction and the area of atelectasis as measured by computed
tomography during anesthesia (shunt = 1.6 × atelectatic area + 1.7) (7). If such results apply to this experiment,
bv would
have resulted in a net 3% greater atelectatic area recruited
than with
c. Although of small magnitude, this change would
represent 3 to 4 times greater area if such lung were aerated
(8) (in this instance a difference in area of 9 to 12%). At 5 h,
S/
T was returned to baseline with
bv but remained elevated at almost 160% of normal with
c and 330% of normal
with
s, suggesting near complete recruitment of atelectatic
lung with "noisy" ventilation. The lower PaCO2 in the
bv
group also suggests better matching of ventilation to perfusion
(
A/
).
Examination of Figure 1 reveals 4 low-frequency rates in
the modulation file/cycle with instantaneous f approximately
7/min. As mean V T in this group was 14.7 ml/kg, the calculated
volume of the lowest breath rate would be (13.9/7 × 14.7) ml/
kg or 29 ml/kg. Concern was raised that the large VT with the
low-frequency breaths represented sighs and that volume recruitment by this mechanism alone could account for the documented improvement in gas exchange during
bv.
To address this issue, we examined eight additional animals
(
s group), ventilated in the same manner as the original
c group, but submitted to programmed "sighs" of identical magnitude and frequency as the low-frequency, large VT breaths in
the
bv group. Thus, this group of animals was exposed to the
same inflation stress as the
bv group but without the biologically variable noise of the
bv group. Despite essentially identical measurements of gas exchange and respiratory mechanics
at baseline values and during one-lung ventilation,
s was not
associated with any improvement in these parameters as seen
with
bv after 5 h of mechanical ventilation and conferred no
advantage compared with
c alone (see Table 1). PaO2 and
shunt fraction were, in fact, worse with this ventilatory mode,
with the proviso that this was a post hoc comparison.
Although the issue of sighs as an effective volume recruitment maneuver during prolonged ventilation is controversial,
examination of the literature suggests that sighs are only effective when administered as a sustained inflation with high pressures under specific circumstances. Sighs of the magnitude
seen during
bv have not been shown to produce beneficial
effects on Crs or gas exchange (9). Balsys and coworkers have
shown that even larger sighs of 46 ml/kg resulted in unsustained increases in compliance and insignificant increases in
PaO2 in healthy lungs in mechanically ventilated dogs (10).
Bond and coworkers demonstrated that sustained inflation increased respiratory compliance only during conventional mechanical ventilation using low VT (7 ml/kg) and low end-expiratory pressure. Sustained inflation was of no benefit during
conventional ventilation with high VT (14 to 17 ml/kg)
the
range of the current study
or with low VT and high end-expiratory pressure (11). Tusman and coworkers showed that a recruitment strategy (VT up to 18 ml/kg coupled with increasing
levels of positive end-expiratory pressure [PEEP] up to 15 cm
H2O) can improve arterial oxygenation after 40 min of anesthesia (12). Pelosi and coworkers demonstrated that gas exchange and respiratory mechanics were improved with 3 consecutive sighs/min at 45 cm H2O plateau pressure over 1 h in
patients with acute respiratory distress syndrome (ARDS)
ventilated with a lung-protective strategy. The improvements
seen were lost within 1 h after return of ventilatory parameters
to baseline values (13). Thus, improvement in gas exchange
occurs at the expense of increases in mean and Ppaw with sustained alveolar recruitment and it is not surprising that the relatively modest sighs delivered during
s resulted in no benefit
in the current study.
Biologically variable ventilation confers the advantage of
improved gas exchange at an unchanged Paw and a lower
Ppaw than either
c or
s. Exclusive to
bv, the animals also
received many small VT/cycle with the potential for alveolar derecruitment.
Ongoing atelectasis is of significant concern during general
anesthesia with mechanical ventilation (8). The near complete recovery of PaO2 and
S/
T to baseline values with
bv indicates better
A/
matching over time during a lengthy period of
anesthesia in the present study. As such,
bv may be of clinical
relevance for control mode ventilation during anesthesia (14).
All variability files used to date in the laboratory have
demonstrated inverse power law frequency distributions and
all have been obtained by lengthy collections of physiological
signals such as heart rate (1), respiratory rate, and blood pressure (15). Variability in these physiological signals is ubiquitous in mammals (16). Further clarification is necessary to
determine if biological variability is representative of stochastic resonance as defined by Suki and coworkers (2). If such
turns out to be the case, then biological variability to recreate
normal variation in VT and f may be an example of tuned
noise to enhance an output (PaO2) in a nonlinear system (4).
We have demonstrated that the programmed variability follows an inverse power law frequency distribution. This
"noisy" behavior may explain the effectiveness of
bv. It is
important to realize that signals with inverse power law frequency distributions are not random (a = 0 or white noise).
Others have suggested that such biologically variable noise
demonstrates deterministic behavior (19, 20). This experiment provides strong support for the theoretical proposal of how
noisy Ppaw can increase recruitment of collapsed alveoli (2).
It is uncertain if
= 0.11 has optimized the benefits that can
be obtained with
bv in this context. Whether or not such a
"noisy" mechanical ventilator has clinical utility must await
further study. However, it is entirely possible that clinical life
support systems may be improved by programming them for
biologically variable or natural noise.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to W. A. C. Mutch, M.D., Department of Anaesthesia, St. Boniface General Hospital, 409 Taché Avenue, Winnipeg, MB, R2H 2A6 Canada. E-mail: amutch{at}ms.umanitoba.ca
(Received in original form March 25, 1999 and in revised form October 28, 1999).
Some of the concepts discussed in this article are protected by U.S. Patent 5,647,350, "Control of Life Support Systems," owned by Biovar Life Support Inc., jointly held by Drs. W. A. C. Mutch, G. R. Lefevre, the University of Manitoba, and the Crocus Investment Fund.Acknowledgments: The authors thank Barb Robson and Carolyn Gibbs for excellent technical assistance and Mary Cheang (M.Math) for statistical analysis.
Supported by the Crocus Investment Fund and the Industrial Research Assistance Program.
| |
References |
|---|
|
|
|---|
1. Lefevre, G. R., S. E. Kowalski, L. G. Girling, D. B. Thiessen, and W. A. C. Mutch. 1996. Improved arterial oxygenation after oleic acid lung injury in the pig using a computer-controlled mechanical ventilator. Am. J. Respir. Crit. Care Med. 154: 1567-1572 [Abstract].
2. Suki, B., A. M. Alencar, M. K. Sujeer, K. R. Lutchen, J. J. Collins, J. S. Andrade Jr., E. P. Ingenito, S. Zapperi, and H. E. Stanley. 1998. Life-support system benefits from noise. Nature 393: 127-128 [Medline].
3. Suki, B., A. Barabási, Z. Hantos, F. Peták, and H. E. Stanley. 1994. Avalanches and power-law behaviour in lung inflation. Nature 368: 615-618 [Medline].
4. Wiesenfeld, K., and F. Moss. 1995. Stochastic resonance and the benefits of noise: from ice ages to crayfish and SQUIDs. Nature 373: 33-36 [Medline].
5. Suki, B., J. S. Andrade Jr., M. F. Coughlin, D. Stamenovic, H. E. Stanley, M. Sujeer, and S. Zapperi. 1998. Mathematical modeling of the first inflation of degassed lungs. Ann. Biomed. Eng. 26: 608-617 [Medline].
6. West, B. J., and M. Shlesinger. 1990. The noise in natural phenomena. Am. Sci. 78: 40-45 .
7.
Gunnarsson, L.,
L. Tokics,
H. Gustavsson, and
G. Hedenstierna.
1991.
Influence of age on atelectasis formation and gas exchange impairment during general anaesthesia.
Br. J. Anaesth.
66:
423-432
8. Hedenstierna, G. 1998. Gas exchange pathophysiology during anesthesia. In P. H. Breen, editor. Anesthesiology Clinics of North America. W.B. Saunders, Philadelphia. 113-127.
9. Housley, E., N. Louzada, and M. R. Becklake. 1970. To sigh or not to sigh. Am. Rev. Respir. Dis. 101: 611-614 [Medline].
10. Balsys, A. J., R. L. Jones, S. F. P. Man, and A. Wells. 1980. Effects of sighs and different tidal volumes on compliance, functional residual capacity and arterial oxygen tension in normal and hypoxemic dogs. Crit. Care Med. 8: 641-645 [Medline].
11. Bond, D. M., J. McAloon, and A. B. Froese. 1994. Sustained inflations improve respiratory compliance during high-frequency oscillatory ventilation but not during large tidal volume positive-pressure ventilation in rabbits. Crit. Care Med. 22: 1269-1277 [Medline].
12. Tusman, G., S. H. Böhm, G. F. Vazquez de Anda, J. L. do Campo, and B. Lachmann. 1999. "Alveolar recruitment strategy" improves arterial oxygenation during general anaesthesia. Br. J. Anaesth. 82:8-13.
13.
Pelosi, P.,
P. Cadringher,
N. Bottino,
M. Panigada,
F. Carrieri,
E. Riva,
A. Lissoni, and
L. Gattinoni.
1999.
Sigh in acute respiratory distress
syndrome.
Am. J. Respir. Crit. Care Med.
159:
872-880
14.
Mutch, W. A. C.,
G. M. Eschun,
S. E. Kowalski,
M. R. Graham,
L. G. Girling, and
G. R. Lefevre.
2000.
Biologically variable ventilation prevents deterioration of gas exchange during prolonged anaesthesia.
Br.
J. Anaesth.
84:
197-203
15.
Mutch, W. A. C.,
G. R. Lefevre,
D. B. Thiessen,
L. G. Girling, and
R. K. Warrian.
1998.
Computer-controlled cardiopulmonary bypass increases jugular venous oxygen saturation during rewarming.
Ann.
Thorac. Surg.
65:
59-65
16. Baselli, G., S. Cerutti, F. Badilini, L. Biancardi, A. Porta, M. Pagani, F. Lombardi, O. Rimoldi, R. Furlan, and A. Malliani. 1994. Model for the assessment of heart period and arterial pressure variability interactions and of respiration influences. Med. Biol. Eng. Comput. 32: 143-152 [Medline].
17.
Pagani, M.,
F. Lombardi,
S. Guzzetti,
O. Rimoldi,
R. Furlan,
P. Pizzinelli,
G. Sandrone,
G. Malfatto,
S. Dell'Orto,
E. Piccaluga,
M. Turiel,
G. Baselli,
S. Cerutti, and
A. Malliani.
1986.
Power spectral analysis of heart rate and arterial pressure variabilities as a marker of
sympatho-vagal interaction in man and conscious dog.
Circ. Res.
59:
178-193
18.
Rimoldi, O.,
S. Pierini,
A. Ferrari,
S. Cerutti,
M. Pagani, and
A. Malliani.
1990.
Analysis of short-term oscillations of R-R and arterial pressure
in conscious dogs.
Am. J. Physiol.
258:
H967-H976
19. West, B. J., and A. L. Goldberger. 1987. Physiology in fractal dimensions. Am. Sci. 75: 354-365 .
20. Goldberger, A. L., and B. J. West. 1987. Fractals in physiology and medicine. Yale J. Biol. Med. 60: 421-435 [Medline].
This article has been cited by other articles:
![]() |
G. L. Gebber and S. M. Barman Variable rate ventilation strategies for the injured lung/Strategies de ventilation a debit variable pour le poumon blesse Can J Anesth, September 1, 2008; 55(9): 572 - 576. [Full Text] [PDF] |
||||
![]() |
K. F. Froehlich, M. R. Graham, T. G. Buchman, L. G. Girling, N. Scafetta, B. J. West, E. K.-Y. Walker, B. M. McManus, and W. A. C. Mutch Physiological noise versus white noise to drive a variable ventilator in a porcine model of lung injury: [Bruit physiologique versus bruit blanc pour entrainer un respirateur en mode variable dans un modele porcin de lesion pulmonaire] Can J Anesth, September 1, 2008; 55(9): 577 - 586. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Thammanomai, L. E. Hueser, A. Majumdar, E. Bartolak-Suki, and B. Suki Design of a new variable-ventilation method optimized for lung recruitment in mice J Appl Physiol, May 1, 2008; 104(5): 1329 - 1340. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Pillow, N. Hillman, T. J. M. Moss, G. Polglase, G. Bold, C. Beaumont, M. Ikegami, and A. H. Jobe Bubble Continuous Positive Airway Pressure Enhances Lung Volume and Gas Exchange in Preterm Lambs Am. J. Respir. Crit. Care Med., July 1, 2007; 176(1): 63 - 69. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Fisher and S. S. Margulies Modeling the effect of stretch and plasma membrane tension on Na+-K+-ATPase activity in alveolar epithelial cells Am J Physiol Lung Cell Mol Physiol, January 1, 2007; 292(1): L40 - L53. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Goldberger Giles F. Filley Lecture. Complex Systems Proceedings of the ATS, August 1, 2006; 3(6): 467 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Mols, H.-J. Priebe, and J. Guttmann Alveolar recruitment in acute lung injury Br. J. Anaesth., February 1, 2006; 96(2): 156 - 166. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. Arold, B. Suki, A. M. Alencar, K. R. Lutchen, and E. P. Ingenito Variable ventilation induces endogenous surfactant release in normal guinea pigs Am J Physiol Lung Cell Mol Physiol, August 1, 2003; 285(2): L370 - L375. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Suki Fluctuations and Power Laws in Pulmonary Physiology Am. J. Respir. Crit. Care Med., July 15, 2002; 166(2): 133 - 137. [Full Text] [PDF] |
||||
![]() |
A. BOKER, M. RUTH GRAHAM, K. R. WALLEY, B. M. MCMANUS, L. G. GIRLING, E. WALKER, G. R. LEFEVRE, and W.A. C. MUTCH Improved Arterial Oxygenation with Biologically Variable or Fractal Ventilation Using Low Tidal Volumes in a Porcine Model of Acute Respiratory Distress Syndrome Am. J. Respir. Crit. Care Med., February 15, 2002; 165(4): 456 - 462. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Graham, R. K. Warrian, L. G. Girling, L. Doiron, G. R. Lefevre, M. Cheang, and W. A. C. Mutch Fractal or biologically variable delivery of cardioplegic solution prevents diastolic dysfunction after cardiopulmonary bypass J. Thorac. Cardiovasc. Surg., January 1, 2002; 123(1): 63 - 71. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. HOTCHKISS JR., A. B. ADAMS, M. K. STONE, D. J. DRIES, J. J. MARINI, and P. S. CROOKE Oscillations and Noise . Inherent Instability of Pressure Support Ventilation? Am. J. Respir. Crit. Care Med., January 1, 2002; 165(1): 47 - 53. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. TOBIN Critical Care Medicine in AJRCCM 2000 Am. J. Respir. Crit. Care Med., October 15, 2001; 164(8): 1347 - 1361. [Full Text] [PDF] |
||||
![]() |
W. A. C. Mutch, G. R. Lefevre, M. S. Cheang, B. A. Simon, R. G. Brower, H. E. Fessler, C. A. Rohde, and A. J. Nam BIOLOGIC VARIABILITY IN MECHANICAL VENTILATION IN A CANINE OLEIC ACID LUNG INJURY MODEL Am. J. Respir. Crit. Care Med., June 1, 2001; 163(7): 1756 - 1757. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |