|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |
ABSTRACT |
|---|
|
|
|---|
The ability of nebulizers to deliver dextran (nominal molecular mass, 4,000 g/mol) to the lung as an
inhaled aerosol is evaluated by in vitro experimental methods and mathematical models. Dextran in
isotonic saline was aerosolized by four nebulizer types (Pari LC STAR, Hudson T-Updraft II, Acorn II,
and Sonix 2000) at dextran concentrations
400 mg/ml and with 2.5- and 4-ml volume fills. Aerosols inhaled during breath simulation were characterized by in-line phase Doppler anemometry, filter collection, osmometry, and gravimetry. Mathematical models were used to estimate amounts of
the characterized aerosols depositing in the different regions of lung models, and mathematical
models of mucous thickness were then developed to estimate initial concentrations of the depositing dextran in the mucus of each conducting airway generation. Models of three subjects (4 yr old, 8 yr
old, and adult) were used. The high viscosity of the dextran solutions tested (up to seven times that
of water) negatively impacts nebulization, and results in poor performance with most delivery systems tested. Our results suggest that airway mucosal dextran concentrations associated with efficacy
in previous animal and in vitro models are achievable with reasonable delivery times (
12 min) with
only one of the delivery systems/formulations tested: the Pari LC STAR nebulizer, using a 2.5-ml volume fill and a dextran concentration of 200 mg/ml. Finlay WH, Lange CF, King M, Speert DP. Lung
delivery of aerosolized dextran.
| |
INTRODUCTION |
|---|
|
|
|---|
Dextran is an oligosaccharide being considered for use in the treatment of cystic fibrosis (CF) because of the therapeutic potential it has demonstrated in animal and in vitro experiments. For example, Feng and coworkers (1) show that dextran exhibits significant mucolytic activity in vitro in CF sputum, while Feng and colleagues (2) show enhanced mucociliary clearance rates when aerosolized dextran is delivered in dogs and suggest that dextran reduces cross-linkage bonding in the mucus, leading to reduced mucous viscoelastic modulus. In addition, Barghouthi and colleagues (3) find that dextran interferes with bacterial adhesion to epithelial cells, and Bryan and coworkers (4) find that aerosolized dextran prevents Pseudomonas aeruginosa pneumonia and death in neonatal mice. For these reasons, dextran may be useful in the treatment of the respiratory aspects of cystic fibrosis, where impaired mucociliary clearance is a major cause of the bacterial infections that lead to much of the morbidity associated with this disease.
However, before the therapeutic benefits of dextran can be examined in human clinical trials, an appropriate delivery system is required. Because dextran appears to demonstrate efficacy in the above-mentioned models owing to local activity in the airway surface fluid, the localized delivery of dextran to airway surfaces is expected to be an effective choice of delivery route. A standard approach for such delivery is the use of an inhaled aerosol. Because nebulizers are uniquely suited to deliver the relatively large quantities of aerosolized dextran that are expected to be needed for in vivo efficacy, nebulization of aqueous dextran is a logical choice of delivery method.
In the present work, we examine the potential of several nebulizer systems to deliver dextran as an inhaled aerosol. The data we present are used to suggest an optimized delivery system that is expected to successfully deliver appropriate amounts of dextran in planned clinical trials.
| |
METHODS |
|---|
|
|
|---|
Dextran 4000 (nominal molar mass, 4,000 g/mol) supplied by Dextran Products (Scarborough, ON, Canada) was prepared in isotonic saline (0.9%, wt/vol) and pH adjusted to 7.4 by addition of 0.1 N NaOH.
Surface tension measurements of the dextran-saline solutions were made by axisymmetric drop shape analysis (5) at room temperature (23 ± 1° C). Measurements of viscosity were made with a rotating spindle viscometer (Brookfield Engineering Laboratories, Stoughton, MA) at 23 ± 1° C room temperature and at 18 ± 1° C, which is the range of temperatures measured during nebulization of dextran with jet nebulizers. Measurements of surface tension were done in triplicate. Ten measurements of viscosity were made at dextran concentrations of 50 and 100 mg/ml, while 15 were made at concentrations of 200 and 400 mg/ml.
Nebulizers were filled with either 2.5 or 4 ml of dextran solutions with differing dextran concentrations. Four different nebulizer types were used. These included two of the most commonly used nebulizers in CF therapy (6), as well as two more recently developed nebulizers that are known to outperform many nebulizers with other formulations (7, 8). The nebulizers included three jet nebulizers: the LC STAR (Pari, Starnberg, Germany), the T-Updraft II (model 1732; Hudson RCI, Temecula, CA), and the Acorn II (Marquest Medical Products, Englewood, CO), in addition to one type of ultrasonic nebulizer, the Sonix 2000 (model 86111; Medix, Catthorpe, UK). A Pulmo-Aide compressor (model 5650D; DeVilbiss, Somerset, PA) was used to supply compressed air to the nebulizers. Flow rates of air supplied by the compressor were measured with two of each of the jet nebulizer types using a dry gas meter (DTM-115; American Meter Division, Horsham, PA). All nebulizers were operated under measured room conditions of 50 ± 10% RH and 22 ± 2° C. Unless otherwise stated, 5 units of each nebulizer type were tested under each set of experimental conditions, and manufacturer lot 1352 of dextran was used. Nebulization was stopped when there was a pause of more than 15 s without aerosol production.
Each nebulizer was tested using procedures described previously by Finlay and coworkers (7), and outlined briefly here. The aerosol path and measurement procedure were designed so that the measured aerosol properties were unaffected by the measurement process. For the LC STAR and Sonix 2000 nebulizers, this involved connecting a breath simulator to the mouthpiece of the nebulizer and collection of "inhaled" aerosol onto a spun polypropylene filter made from 3M Filtrete (Marquest No. 303; Marquest Medical Products). For the Acorn II and T-Updraft II nebulizers, amounts of inhaled aerosol were inferred directly from total amounts of dextran nebulized, using the inhalation:exhalation ratios in the breathing patterns given below. For all nebulizers, initial and final amounts of dextran were determined from pre- and postnebulization measurements of concentration, using freezing point osmometry (Precision Systems, Natick, MA), and of mass.
The breath simulator consisted of an in-house computer-controlled piston and was used with the following tidal breathing patterns. An adult cystic fibrosis pattern was developed on the basis of data in the literature on breathing patterns of patients with cystic fibrosis (9). This pattern consisted of a tidal volume of 0.62 L, inhalation:exhalation time ratio of 1:1.3, 18 breaths/min, and sine wave shapes for the inhalation and exhalation cycles. Two pediatric breathing patterns were also tested with the LC STAR nebulizer. These simulated a 4-yr-old child (square wave, tidal volume of 0.23 L, inhalation flow rate of 11 L/min), and an 8-yr-old child (square wave, tidal volume of 0.32 L, inhalation flow rate of 13 L/min), based on data in the literature (15- 17). A second adult breathing pattern consisting of a square wave with tidal volume of 0.75 L, 12 breaths/min (inhalation flow rate of 18 L/ min) was used for interlot testing of dextran with the LC STAR nebulizer and dextran manufacturer lots 1352, 2750, 2758, and 2764.
Droplet size measurements of the aerosols exiting the nebulizers were done using phase Doppler anemometry (Dantec Electronics, Mahwah, NJ). For the two vented nebulizers (LC STAR and Sonix 2000), droplet size measurements were made during the simulated tidal breathing by using an enclosed, in-line optically clear sizing region consisting of a cylinder with flat optical windows, as described by Prokop and coworkers (18). For the two unvented nebulizers (Acorn II and T-Updraft II), droplet size measurements were made using the procedure of Stapleton and colleagues (19) by removing the T-mouthpieces and measuring the aerosol exiting the nebulizer prior to exposure to any ambient air.
Estimates of regional lung deposition of the aerosol inhaled from
the LC STAR were made using a mathematical lung deposition model
similar to that described by Finlay and colleagues (7), which compares
well with in vivo
scintigraphic measurements on normal subjects (20,
21). In this model, a symmetrically branching model of the lung is
used based on the analysis of morphometric data given by Phillips and
coworkers (22) for the conducting airways and by Haefeli-Bleuer and
Weibel (23) for the alveolar region. Denoting the trachea as generation 0, the conducting airways of this lung model occupy generations
0-14, and the alveolar region occupies generations 15-23. The lung
model was scaled to give pediatric lung models for children, 4 and 8 yr
old, using the procedures of Hofmann and colleagues (24). With this
procedure, the 4-yr-old lung model has one less generation of alveoli
than the adult and 8-yr-old lung models. The diameters and lengths of
the generations of the different lung models are given in Table 1.
Lung volumes of the three models are 3,000 ml for the adult, 1,135 ml
for the 8-yr-old, and 529 ml for the 4-yr-old. Amounts of aerosol depositing in each generation of these lungs with the above-described
breathing patterns were estimated using a one-dimensional Lagrangian
approach and the equations of Chan and Lippmann (25) for inertial
impaction, those of Pich (26) and Heyder and Gebhart (27) for sedimentation, and those of Gormley and Kennedy (28) for diffusion.
Deposition in the mouth-throat was estimated using the equations of
Rudolf and colleagues (29). Hygroscopic effects were not included because of the high aerosol mass fraction produced by the LC STAR,
which is sufficient to eliminate such effects. Indeed, the parameter
proposed by Finlay (30) to estimate the importance of such effects
was well within the range where hygroscopic effects are negligible
(
3), and simulations done with inclusion of two-way coupled hygroscopic effects differed by < 3% in predictions of regional dosages
from values obtained with such effects neglected.
|
Estimates of the average concentration of dextran in the mucus
immediately after completion of aerosol delivery were made by combining the above-described deposition estimates with estimates of mucous volume within each airway generation, assuming uniform disposition of depositing dextran and mucus within each generation. Mucous
volume in each lung generation was determined from mucous thickness estimates with the airway lengths and diameters given in Table 1.
Mucous thickness was estimated using mass conservation and a model
of generational mucous velocities for which values of the tracheal mucous velocity and the mucous volume production rate were specified.
For the adult model, tracheal mucous velocities of 5 and 15 mm/min
and a mucous production rate of 10 ml/d were used, which are representative of values in the literature (31). For the two pediatric
ages, tracheal mucous velocities of 5 and 15 mm/min were used with
mucous production rates obtained by scaling the adult value by body
weight (giving 2.6 and 4 ml/d for the 4- and 8-yr-old child models, respectively). Mucous velocities were estimated using a procedure similar to that described by Cuddihy and Yeh (34), Yu and colleagues
(35), and Yu (36). With this procedure, the mucociliary clearance of
the tracheobronchial region is treated as a series of "escalators." Using a deposition model to estimate amounts depositing in each generation, the mucous velocities, assumed constant in each generation, are
estimated in order to match the clearance rates of in vivo deposition
and clearance studies. Here we used our above-described regional
deposition model to estimate generational deposition of the various
radiolabeled aerosols inhaled in the in vivo study of Stahlhofen and
coworkers (37). Generational mucous velocities were then determined in order to match the average in vivo clearance rates given by
Stahlhofen and coworkers (37). It is possible to estimate the time
needed to clear the kth generation, for example, by taking the deposited amount predicted by the deposition model for this generation and using the clearance curve to obtain the time needed to clear the kth
fast-cleared fraction corresponding to the same amount. The generational mucous velocity is then estimated as Uk = Lk/
k, where Lk is the
length of the kth generation and
k is the time needed to clear this
generation. The mucous thickness follows by calculating the annular
cross-section that corresponds to the same mucous volume flow rate
as prescribed at the trachea. The mucous thickness of the most distal
tracheobronchial generation (generation 14) was set equal to that estimated for generation 13.
Statistical analysis of the results was done using ANOVA (SYSTAT; SYSTAT, Evanston, IL), with Tukey highest significant differences (HSD) multiple means tests. Differences were deemed significant at a p value of 0.01, unless otherwise stated.
| |
RESULTS |
|---|
|
|
|---|
Values of viscosity and surface tension for different dextran concentrations are shown in Figures 1 and 2. Surface tension varies only marginally with dextran concentration (varying by < 6%) and differs little from that of water (73 mN/m), although these differences are significant (p < 0.01). The viscosity of the dextran solutions shown in Figure 2 is significantly greater than that of water (1.0 mPa · s). Viscosity varies dramatically with concentration over the range tested, and these variations are significant (p < 0.01). Viscosity is also significantly higher at 18° C than at 23° C (Student t test: p < 0.01).
|
|
Flow rates from the compressor were 5.6 L/min for the Acorn II, 5.7 L/min for the T-Updraft II, and 4.7 L/min for the LC STAR.
Amounts of dextran inhaled, nebulization times, aerosol mass median diameter (MMD), and geometric standard deviation (GSD) are shown at various dextran concentrations in Table 2 for a nebulizer volume fill of 2.5 ml with the adult CF breathing pattern. Similar data are shown in Table 3 for a nebulizer volume fill of 4 ml. Data for the Sonix 2000 is not included in Table 3 because two of the Sonix nebulizers became inoperational after nebulization of 200 mg/ml with the 2.5-ml volume fill, and further testing with these units was not done to avoid possible breakdown of additional units at these high dextran concentrations.
|
|
At each dextran concentration and nebulizer fill volume in Tables 2 and 3, the nebulization times, MMD, and GSD of the LC STAR are less than for the other nebulizers. These differences are significant (p < 0.01).
Average nebulization times for all devices at both volume fills are significantly longer (p < 0.01) at 400 mg/ml than at either 100 or 200 mg/ml, but do not differ significantly between 100 and 200 mg/ml. Nebulization times are also significantly longer (p < 0.01) for all delivery systems at all concentrations when using the 4-ml volume fill compared with the 2.5-ml volume fill.
Mass median diameters and geometric standard deviations were not significantly affected by either volume fill or dextran concentration except for the LC STAR, where MMDs were significantly smaller at 400 mg/ml (p < 0.01), and the T-Updraft II, where GSDs were significantly larger at 400 mg/ml with a 4-ml volume fill.
The amount of dextran inhaled as a percentage of that placed in the delivery device decreased significantly between 200 and 400 mg/ml (p < 0.05), but did not differ significantly between 100 and 200 mg/ml except for the Sonix 2000 (where significantly less was delivered at 200 mg/ml, and data were not taken at 400 mg/ml). At each dextran concentration, the percentages of inhaled dextran for the three jet nebulizers were not significantly affected by volume fill. Although significant differences exist in amounts of inhaled dextran between different nebulizer types, these differences are not significant among the three jet nebulizer types for a 2.5-ml volume fill at dextran concentrations of 100 and 200 mg/ml.
Interlot variations in the amounts of dextran inhaled (0.75-L
tidal volume, 18-L/min inhalation/exhalation flow rate) with four different dextran lots are shown in Figure 3 for the LC
STAR, using a 2.5-ml volume fill and various dextran concentrations. Interlot differences are not significant (p > 0.01) at
each concentration
200 mg/ml shown in Figure 3, but are
significant at 300 mg/ml.
|
Results from further testing of LC STAR nebulizers with the 4-yr-old and 8-yr-old breathing patterns are shown in Table 4 (using dextran lot 1352) at a dextran concentration of 200 mg/ml. Data from Table 2 for the adult case are repeated here for the reader's convenience. Mass median diameters did not differ significantly between the three ages (p > 0.05). Nebulization times are significantly longer for the 4-yr-old breathing pattern (p < 0.05), but did not differ significantly between the two older age breathing patterns. Amounts of inhaled dextran increased significantly with age (p < 0.01).
|
Table 4 also shows amounts of dextran predicted to deposit in the different regions of the mathematical lung models. Estimated amounts depositing in the alveolar and tracheobronchial regions of the mathematical lung models varied significantly with age (p < 0.01), but amounts in the extrathoracic (mouth-throat) region did not (p > 0.05).
Concentrations of dextran in the mucus immediately after aerosol exposure as estimated with our mucous model are shown in Figure 4 for the two assumed tracheal mucous velocities (5 and 15 mm/min) and the assumed mucous production rates given earlier in METHODS. Error bars are not shown, because they are smaller than the symbol sizes in Figure 4. The estimated mucous concentrations increase significantly with age and decrease significantly with mucous velocity (p < 0.01). Mucous concentrations also decreased in nearly direct proportion with mucous production rate. For example, increasing mucous production rate by a factor of 10 reduced dextran concentrations on average by a factor of 9.6 in the adult model and 9.3 in the 8-yr-old model.
|
| |
DISCUSSION |
|---|
|
|
|---|
Our results show that as dextran concentrations increase above 200 mg/ml, nebulization is adversely affected. In particular, nebulization times increase dramatically (e.g., almost doubling between 200 and 400 mg/ml for the T-Updraft II and a 4-ml volume fill), nebulizer efficiencies decrease, where nebulizer efficiency is defined as the amount of inhaled dextran as a percentage of that placed in the nebulizer (e.g., the Acorn II delivers almost four times as much dextran at 200 mg/ml compared with 400 mg/ml), and interlot variations in dextran formulation begin to cause significant interlot differences in amounts of inhaled dextran. For these reasons, the use of dextran concentrations greater than ~ 200 mg/ml for nebulizer delivery does not appear to be warranted.
The dramatic adverse changes in nebulizer behavior that occur at dextran concentrations greater than 200 mg/ml appear to be the result of the rapid increase in viscosity of the dextran solutions that occurs at these concentrations, since surface tension varies only slightly with dextran concentration, while the viscosity nearly triples from 200 to 400 mg/ml to a value that is six to seven times that of water, as seen in Figures 1 and 2. Failure of ultrasonic nebulizers to operate with such viscous liquids has been observed by other authors (see Reference 38), and the increase in viscosity at the higher concentrations is the probable explanation for the failure of the Sonix nebulizers.
Viscous effects also appear to be responsible for the differences in amounts of dextran inhaled from the LC STAR at the
three different ages. In particular, the design of this breath-
enhanced nebulizer is such that the lower minute volumes at the
younger ages lead to lower aerosol output rates. This causes
nebulization times to increase at the younger ages. This in turn
leads to enhancement of the concentration effect owing to
evaporation seen with jet nebulizers (39), so that higher dextran concentrations occur over more of the nebulization time
with the younger ages. This results in more viscous droplets
because of the effect of dextran concentration on viscosity (see
Figure 2), leading to increased holdup of drops on the nebulizer walls and baffles due to decreased ability of droplets to
flow back into the nebulizer solution. Increased holdup of
dextran on the interior walls of the device results in increased
wastage of dextran at the end of nebulization, and less delivered as aerosol, as indeed seen in Table 4. This explanation
also implies that the absence of increases in viscosity with increases in concentration should remove this effect. This is observed to be the case, since data we obtained using isotonic saline alone (n = 3) showed no significant decreases (p > 0.01) in amounts of inhaled salt between the adult CF breathing
pattern and either child breathing pattern with the LC STAR
and a 2.5-ml volume fill. Indeed, average inhaled amounts
with the LC STAR were significantly larger for isotonic saline
alone than with dextran (as a percentage of that placed in the
nebulizer, averaging
30% with isotonic saline alone), also
because of the described viscous effects.
A similar effect involving increased holdup due to viscosity
can be used to explain the lack of expected improvement in
nebulizer efficiency with 4- versus 2.5-ml volume fills. Because
nebulization times are considerably longer (
16 min) when
volume fills of 4 ml were used instead of 2.5 ml, and because
nebulizer efficiency was not significantly different between
the two volume fills, a volume fill of 2.5 ml is preferable in order to give reasonable expectations of patient compliance.
Because the LC STAR was the only nebulizer that gave
nebulization times
12 min at all concentrations up to 200 mg/ml with a 2.5-ml volume fill, and since none of the other
nebulizers delivered significantly more inhaled dextran than
the LC STAR with this volume fill at concentrations up to 200 mg/ml, the LC STAR is the logical choice of delivery system
for the dextran formulations considered here.
The in vitro experiments of Barghouthi and coworkers (3)
suggest that dextran mucous concentrations of approximately
40 mg/ml are optimal in reducing bacterial adhesion, although significant reductions in adhesion are observed at dextran concentrations as low as a few milligrams per milliliter.
Furthermore, the in vitro studies of Feng and colleagues (1)
indicate that effective mucolysis (significant reduction in mucous viscoelastic modulus) occurs in the concentration range
4-40 mg/ml. Using the LC STAR nebulizer at dextran concentrations of 200 mg/ml and a 2.5-ml volume fill, the results from
our mathematical models (Figure 4) show that mucous concentrations
40 mg/ml occur throughout most of the tracheobronchial airways in our 4-yr-old lung model. Dextran mucous
concentrations less than 40 mg/ml are predicted in our 8-yr-old and adult lung models with this delivery system, although
concentrations
4 mg/ml are predicted throughout much of
the conducting airways, so that some antiadhesive and mucolytic effects might still be expected at these two older ages. Increased mucous production rates would be expected to lead to
decreased mucosal dextran concentrations, as indeed is observed in the results with our model, but even a factor of 10 increase in mucous production rate gives concentrations
4 mg/ml throughout much of the 4-yr-old lung model. If higher
concentrations than those obtained with a 2.5-ml volume fill
are desired, our results with 4-ml fills indicate that the use of
larger volume fills should result in nearly directly proportional
larger mucosal concentrations, provided the longer nebulization times are acceptable.
It should be noted that our mucous model provides an estimate of mucous concentrations only immediately after aerosol treatment, since it does not include the effect of clearance or absorption on dextran mucous concentrations. However, neglecting clearance and absorption is expected to be reasonable in estimating initial dextran mucous concentrations, since the time constants for clearance and absorption (32, 40) of the dextran used here (nominal molar mass, 4,000 g/mol) are expected to be much longer than the aerosol delivery times (10 min) for the LC STAR nebulizer with which this model was used. The model may also be useful in aiding delivery system design for other new therapeutic agents, such as antibiotics, where efficacy occurs via immediate onset of local activity in the airway surface fluid.
Because our lung models do not include any morphological or physiological changes in the lung associated with the presence of cystic fibrosis, our model calculations must be interpreted with due care. In particular, increased mucous production rates and slower mucus transport rates will decrease mucosal dextran concentrations, as is clearly seen from our results at the lower tracheal mucous velocities and higher mucous production rates given in Results. However, our results do provide indications that the amounts of dextran delivered to the airways using the LC STAR nebulizer/Pulmo-Aide compressor with dextran at 200 mg/ml and a 2.5-ml volume fill are enough that this delivery system is appropriate for consideration in future clinical trials with patients whose tracheal mucous velocities and mucous production rates are similar to those considered here. Because of concerns that this delivery system may give optimal dextran concentrations only in younger children, consideration of delivery systems that yield higher concentrations of dextran in the mucus (without increasing treatment times) would be worthwhile. Note, however, that this delivery system is already delivering 20-30 mg of dextran to the conducting airways according to our lung models (and total inhaled amounts of dextran near 100 mg). These are doses that are beyond the capabilities of many aerosol delivery systems (such as typical dry powder or pressurized metered dose inhalers).
Summary
The ability of nebulizers to deliver aqueous dextran to the lung as an inhaled aerosol was examined. Experimental methods were used to characterize the aerosols inhaled during simulated tidal breathing. Nebulization deteriorated significantly at dextran concentrations greater than 200 mg/ml, resulting in excessive nebulization times and reductions in the amount of inhaled dextran caused by the high viscosity of these dextran solutions. Volume fills of 4 ml were not associated with higher nebulizer efficiency compared with 2.5 ml, but instead gave excessive nebulization times. At 200 mg/ml and a 2.5-ml volume fill, the Pari LC STAR had significantly shorter nebulization times and smaller droplet sizes than the other delivery systems, while delivering similar amounts of dextran.
Mathematical lung deposition models were combined with
mucus thickness models and the above-described experimental measurements of inhaled aerosols to estimate mucosal dextran concentrations with the Pari LC STAR nebulizer for a 4-yr-old child, an 8-yr-old child, and an adult. Mucosal dextran
concentrations are predicted to be
40 mg/ml with this delivery system in our 4-yr-old model. Such concentrations are associated with optimal positive effects in previous in vitro bacterial adhesion experiments. In our 8-yr-old and adult models,
significantly lower mucosal concentrations are predicted, although these values are
4 mg/ml over much of the airways
and are still associated with significant positive effect in vitro.
Whether these results translate into in vivo efficacy with this
delivery system remains for future clinical trials to determine.
| |
Footnotes |
|---|
Correspondence and requests for reprints should be addressed to Warren H. Finlay, Aerosol Research Laboratory, Department of Mechanical Engineering, University of Alberta, Edmonton, AB, T6G 2G8 Canada. E-mail: warren.finlay{at}ualberta.ca
(Received in original form December 16, 1998 and in revised form June 28, 1999).
Supported financially by the Canadian Cystic Fibrosis Foundation. Dextran was generously supplied as a gift from Dextran Products, Ltd.Acknowledgments: The laboratory help of H. Orszanska is gratefully acknowledged.
| |
References |
|---|
|
|
|---|
1.
Feng, W.,
H. Garret,
D. P. Speert, and
M. King.
1998.
Improved clearability of cystic fibrosis sputum with dextran treatment in vitro.
Am. J. Respir Crit. Care Med.
157:
710-714
2. Feng, W., S. Nakamura, E. Sudo, M. M. Lee, A. Shao, and M. King. 1999. Effects of dextran on tracheal mucociliary velocity in dogs in vivo. Pulm. Pharmacol. Ther. 12: 35-41 . [Medline]
3. Barghouthi, S., L. M. Guerdoud, and D. P. Speert. 1996. Inhibition by dextran of Pseudomonas aeruginosa adherence to epithelial cells. Am. J. Respir. Crit. Care Med. 154: 1788-1793 [Abstract].
4. Bryan, R., M. Feldman, S. C. Jawetz, S. Rajan, E. DiMango, L. Scheffler, D. P. Speert, and A. Prince. 1999. The effects of aerosolized dextran in a mouse model of Pseudomonas aeruginosa pulmonary infection. J. Infect. Dis. 179: 1449-1458 [Medline].
5. Li, D., P. Cheng, and A. W. Neumann. 1992. Contact angle measurement by axisymmetric drop shape analysis (ADSA). Adv. Colloid Interface Sci. 39: 347-382 .
6. Rosenfeld, M., J. Emerson, S. Astley, P. Joy, J. Williams-Warren, T. A. Standaert, D. L. Yim, D. Crist, M. Thykkuttathil, M. Torrence, S. Fitzsimmons, and B. Ramsey. 1998. Home nebulizer use among patients with cystic fibrosis. J. Pediatr. 132: 125-131 [Medline].
7. Finlay, W. H., K. W. Stapleton, and P. Zuberbuhler. 1998. Variations in predicted regional lung deposition of salbutamol sulphate between 19 nebulizer models. J. Aerosol Med. 11: 65-80 .
8. Finlay, W. H., and J. P. Wong. 1998. Regional lung deposition of nebulized liposome-encapsulated ciprofloxacin. Int. J. Pharmaceut. 167: 121-127 .
9.
Diot, P.,
L. Palmer,
A. Smaldone,
J. DeCelie-Germana,
R. Grimson, and
G. C. Smaldone.
1997.
RhDNase I aerosol deposition and related factors in cystic fibrosis.
Am. J. Respir. Crit. Care Med.
156:
1662-1668
10. Ballard, R. D., J. M. Sutarik, C. W. Clover, and B. Y. Suh. 1996. Effects of non-REM sleep on ventilation and respiratory mechanics in adults with cystic fibrosis. Am. J. Respir. Crit. Care Med. 153: 266-271 [Abstract].
11. Denyer, J., A. Dyche, and K. Nikander. 1996. Breathing patterns in adult patients. ISAM Focus Symposium on Aerosol Therapy with Small Volume Nebulizers: Laboratory to Bedside. Tours, France, Sept. 4-5.
12.
Cerny, F.,
L. Armitage,
J. A. Hirsch, and
B. Bishop.
1992.
Respiratory
and abdominal muscle responses to expiratory threshold loading in
cystic fibrosis.
J. Appl. Physiol.
72:
842-850
13.
Browning, I. B.,
G. E. D'Alonzo, and
M. J. Tobin.
1990.
Importance of
respiratory rate as an indicator of respiratory dysfunction in patients
with cystic fibrosis.
Chest
97:
1317-1321
14.
Bureau, M. A.,
L. Lupien, and
R. Bégin.
1981.
Neural drive and ventilatory strategy of breathing in normal children, and in patients with cystic fibrosis and asthma.
Pediatrics
68:
187-194
15. ATS/ERS. 1993. Respiratory mechanics in infants: physiologic evaluation in health and disease, 1993. Am. Rev. Respir. Dis. 147: 474-496 [Medline].
16. Hofmann, W.. 1982. Mathematical model for the postnatal growth of the human lung. Respir. Physiol. 49: 115-129 [Medline].
17. Taussig, L. M., T. R. Harris, and M. D. Lebowitz. 1977. Lung function in infants and young children. Am. Rev. Respir. Dis. 116: 233-239 [Medline].
18. Prokop, R. M., W. H. Finlay, and K. W. Stapleton. 1995. An in vitro technique for calculating the regional dosages of drugs delivered by an ultrasonic nebulizer. J. Aerosol Sci. 26: 847-860 .
19. Stapleton, K., W. H. Finlay, and P. Zuberbuhler. 1994. An in vitro method for measuring regional dosages delivered by jet nebulizers. J. Aerosol Med. 7: 325-344 [Medline].
20. Finlay, W. H., M. Hoskinson, and K. W. Stapleton. 1998. Can models be trusted to subdivide lung deposition into alveolar and tracheobronchial fractions? In R. N. Dalby, P. R. Byron, and S. J. Farr, editors. Respiratory Drug Delivery VI. Interpharm Press, Buffalo Grove, IL. 235-224.
21.
Finlay, W. H.,
K. W. Stapleton,
H. K. Chan,
P. Zuberbuhler, and
I. Gonda.
1996.
Regional deposition of inhaled hygroscopic aerosols: in
vivo SPECT compared with mathematical deposition modeling.
J.
Appl. Physiol.
81:
374-383
22. Phillips, C. G., S. R. Kaye, and R. C. Schroter. 1994. A diameter-based reconstruction of the branching pattern of the human bronchial tree: I. Description and application. Respir. Physiol. 98: 193-217 [Medline].
23. Haefeli-Bleuer, B., and E. R. Weibel. 1988. Morphometry of the human pulmonary acinus. Anat. Rec. 220: 401-414 [Medline].
24. Hofmann, W., T. B. Martonen, and R. C. Graham. 1989. Predicted deposition of nonhygroscopic aerosols in the human lung as a function of subject age. J. Aerosol Med. 2: 49-68 .
25. Chan, T. L., and M. Lippmann. 1980. Experimental measurements and empirical modeling of the regional deposition of inhaled particles in humans. Am. Ind. Hyg. Assoc. J. 41: 399-409 [Medline].
26. Pich, J.. 1972. Theory of gravitational deposition of particles from laminar flows in channels. J. Aerosol Sci. 3: 351-361 .
27. Heyder, J., and J. Gebhart. 1977. Gravitational deposition of particles from laminar aerosol flow through inclined circular tubes. J. Aerosol Sci. 6: 311-328 .
28. Gormley, P. G., and K. Kennedy. 1949. Diffusion from a stream flowing through a cylindrical tube. Proc. R. Irish Soc. 52A:163.
29. Rudolf, G., R. Köbrich, and W. Stahlhofen. 1990. Modeling and algebraic formulation of regional deposition in man. J. Aerosol Sci. 21: S403-S406 .
30. Finlay, W. H.. 1998. Estimating the type of hygroscopic behaviour exhibited by aqueous droplets. J. Aerosol Med. 11: 221-229 [Medline].
31. ICRP (International Commission on Radiological Protection). 1994. Human Respiratory Tract Model for Radiological Protection. Annals of the ICRP 24 (1-3). Elsevier Science, Tarrytown, NY.
32. Wanner, A., M. Salathé, and T. G. O'Riordan. 1996. Mucociliary clearance in the airways. Am. J. Respir. Crit. Care Med. 154: 1868-1902 [Medline].
33. King, M., and B. K. Rubin. 1996. Mucus physiology and pathophysiology: therapeutic aspects. In J. P. Derenee, W. A. Whitelaw, and T. Similowski, editors. Acute Respiratory Failure in COPD (Lung Biology in Health and Disease Series). Marcel Dekker, New York. 391- 411.
34. Cuddihy, R. G., and H. C. Yeh. 1988. Respiratory tract clearance of particles and substances dissociated from particles. In U. Mohr, editor. Inhalation Toxicology: The Design and Interpretation of Inhalation Studies and Their Use in Risk Assessment. Springer-Verlag, Berlin. 169-193.
35. Yu, C. P., J. P. Hu, B. M. Yen, D. M. Spektor, and M. Lippmann. 1986. Models for mucociliary particle clearance in lung airways. In S. D. Lee, T. Schneider, L. D. Grant, and P. J. Verker, editors. Aerosols: Research, Risk Assessment and Control Strategies. Lewis, Chelsea, MI. 569-578
36. Yu, C. P. 1981. A model for particle clearance in human tracheobronchial tree. In Proceedings of 34th ACEMB, Houston, TX, September 21-23, 1981. Alliance for Engineering in Medicine and Biology, Bethesda, MD. 39.
37. Stahlhofen, W., J. Gebhart, and J. Heyder. 1980. Experimental determination of the regional deposition of aerosol particles in the human respiratory tract. Am. Ind. Hyg. Assoc. J. 41: 385-398 [Medline].
38. Taylor, K. M. G., and O. N. M. McCallion. 1997. Ultrasonic nebulizers for pulmonary drug delivery. Int. J. Pharm. 153: 93-104 .
39. Mercer, T. T., M. I. Tillery, and H. Y. Chow. 1968. Operating characteristics of some compressed-air nebulizers. Am. Ind. Hyg. Assoc. J. 29: 66-78 [Medline].
40. Byron, P. R.. 1986. Prediction of drug residence times in regions of the human respiratory tract following aerosol inhalation. J. Pharm. Sci. 75: 433-438 [Medline].
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Proc. Am. Thorac. Soc. | Am. J. Respir. Cell Mol. Biol. |